Rapid Analysis of Hydrocarbon Stream Composition in Spark Ignition Engine Fuels with Concurrent Heating

Topics Covered

Introduction
PIONA+ Analyzer
Instrumentation used
Results and discussion
Conclusion
About Bruker - Chemical and Applied Markets

Introduction

Gas chromatography (GC) is highly versatile and widely used technique for characterizing volatile chemical compounds. It can be used to quantitatively analyze the types of hydrocarbon (olefins, napthalenes, aromatics, etc.) in a sample of hydrocarbons, which makes it very useful for examining fuel compositions.

GC analyzers from Bruker are configured and tested to exceed industry standard performance requirements for a variety of widely used methods (ASTM, UOP, EN, ISO, GPA). Available in an array of configurations, Bruker Standard GC Analyzers utilize the leading analytical performance characteristics of the Bruker 400 Series of Gas Chromatography Systems.

PIONA+™ Analyzer

The PIONA+ Analyzer from Bruker is especially designed and configured to be fully compliant with any of the following established methods including: ASTM D6839, EN-ISO-22854, ASTM D5443, ASTM D1319, IP-382, IP-526, EN 14157, UOP 870 and DIN-51448.

The PIONA+ uses an automated multidimensional GC approach to analyze hydrocarbon content by type and carbon number, which allows complete characterization of a hydrocarbon sample, such as spark ignition engine fuels, including individual oxygenates.

The sample is separated in the component groups per carbon number and in individual components through the use of multiple columns and traps. In the PIONA+ system, paraffins, iso-paraffins and iso-olefins, olefins, naphthenes and aromatics are identified.

However, analysis time in the PIONA mode is about 180 minutes, which severely limits the number of samples that can be analyzed per day. A unique aspect of the design of the Bruker PIONA+ Analyzer is the ability to independently heat the individual traps (concurrent heating). This application note describes work to determine if concurrent heating could be used to significantly reduce the analysis time and improve sample throughput.

Instrumentation used

Results and Discussion

As shown in Figure 1, when the PIONA+ system is operated in the conventional mode, a total analysis time of 180 minutes is required to elute all of the component groups. However, as shown in Figure 2, through the use of concurrent heating of both Bruker Molsieve 5A and 13X traps, a reduction in total analysis time from 180 minutes to 95 minutes is obtained, i.e. ~50 % less.


Figure 1: Chromatogram of a test sample 1 in conventional PIONA mode.

The reduction in analysis time is achieved by carefully optimizing the temperature settings of the various traps and columns and invoking simultaneous heating of the Molsieve traps. This approach, referred to as the “FastPIONA” mode, results in the elution of the paraffins immediately after their naphthene and iso-paraffin counterparts.

Yet the elution integrity of the component groups remains intact with no negative influence on either naphthene or iso-paraffin groups. Figure 3 shows a close-up of the resulting elution sequence. The same results occur for the olefin group separations. The cyclic and iso-olefins normally elute from 100–130 minutes and the n-olefins, in conventional PIONA mode, from 150–180 minutes.

However, through the use of concurrent heating they completely elute between 55 and 70 minutes (Figure 4). In the example of Figure 5, a commercial standard was analyzed for O-PIONA (oxygenates, paraffins, iso-paraffins, olefins, iso-olefins, naphthenes and aromatics) utilizing the concurrent heating approach described above. Note that the oxygenates elute separately from the other component groups and are easily identified and calculated by the PIONA+ software.


Figure 2. Chromatogram of the test sample in fast PIONA mode.


Figure 3. Close-up view of C8 and C9 components. Separation in both conventional and fast
PIONA mode.

Conclusion

Temperature programming a Molsieve 5A trap (concurrent heating) together with a Molsieve 13X trap greatly reduces the total analysis time for hydrocarbon streams with the PIONA+ Analyzer and, in some cases, can improve the performance.

Total analysis time is reduced by one third in the case of O-PIONA and by as much as half for PIONA. The reduction in analysis time is possible without any hardware changes and can be achieved without any negative effect on the quality of the component class separation. In addition, the analyses remain compliant with the method EN ISO 22854:2008.


Figure 4. Close-up view of olefins in conventional and fast PIONA mode.

Figure 5. Chromatogram of a commercial standard in fast O-PIONA mode.

About Bruker - Chemical and Applied Markets

Bruker is the new name in chemical analysis. Accurate and comprehensive analysis of exogenous and discrete elements in a wide range of sample matrices are key applications for many analytical chemistry groups. To address the needs and challenges of analysts working in those areas, Bruker has expanded their product family to provide, and expertly support, a series of fully integrated solutions including:

  • Gas Chromatography-Mass Spectrometers (GC/MS and GC/MS/MS)
  • Inductively Coupled Plasma Mass Spectrometers (ICP-MS)
  • Gas Chromatography Systems (GC)

Widely used in food and consumer safety testing, forensic, industrial, environmental, and clinical laboratories, these systems are well accepted and established market leaders that universally deliver outstanding performance at a premium value.

This information has been sourced, reviewed and adapted from materials provided by Bruker - Chemical and Applied Markets (CAM).

For more information on this source please visit Bruker - Chemical and Applied Markets.

Date Added: Jun 15, 2012 | Updated: Jun 11, 2013
Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Submit