Use of Headspace-Gas Chromatography for Transformer Oil Gas Analysis

Topics Covered

Introduction
Instrumentation Used
Materials and Reagents
Sample Preparation
Conditions
Results and Discussion
Conclusion
About Bruker - Chemical and Applied Markets

Introduction

Insulating fluids, generally mineral oils, are used in transformers. Under normal, mild conditions there is very little decomposition.

Occasionally however (localized or general) overheating of the oil occurs and decomposition products are formed. If the concentration of these gases reaches a critical point, the chances of catastrophic transformer failure are high.

The ASTM D 3612 method describes in detail three different routes:

A. Vacuum Extraction

The gases are extracted from the oil via a vacuum extraction device and analyzed via gas chromatography.

B. Stripper Column Extraction

Dissolved gases are extracted from a sample of oil by sparging the oil with the carrier gas on a stripper column containing a high surface area bead. The gases are then flushed from the stripper column into a gas chromatograph for analysis.

C. Headspace Sampling

An oil sample is brought into contact with a gas phase (headspace) in a closed vessel purged with Argon. As a result, a portion of a gas dissolved in the oil is transferred to the Headspace.

This article describes analysis of a transformer oil by method C - Headspace-Gas Chromatography

Instrumentation Used


Figure 1. TOGA Analysis, TCD Channel

Materials and Reagents

“True North” DGA Oil Standard by Morgan Schaffer:

Hydrogen 88 ppm
Oxygen 11163 ppm
Nitrogen 40368 ppm
Methane 96 ppm
Carbon monoxide
89 ppm
Carbon dioxide
123 ppm
Ethylene 90 ppm
Ethane 92 ppm
Acetylene 84 ppm

Sample Preparation


Figure 2. TOGA analysis FID channel


Figure 3. Schematic overview hardware.

The Morgan Schaffer Calibration Standard is carefully transferred into the headspace vial. The gases are extracted from the oil by means of a headspace sampler and injected onto a short Hayesep P precolumn and then to a micro packed Carboxen-1000 column.

The fraction containing Hydrogen, Oxygen, Nitrogen, Carbon Monoxide, and Methane will elute direct from the Carboxen-1000 column to the micro packed Molsieve column. Hydrogen, Oxygen and Nitrogen are detected by the TCD. Carbon Monoxide, and Methane are detected by the FID, after passing the Methanizer. When the Molsieve column is bypassed, Carbon Dioxide and the C2-C3 isomers are eluting from the Carboxen-1000 column and detected by the FID after passing the Methanizer. The back flush time is set to completely elute the C3 isomers. C4 and higher are back flushed.

Conditions

Table 1. Column oven settings

Rate (°C/min) Step (°C) Time (min.)
Initial 50 5.0
10.0 75 0.0
20.0 220 10.25

Total Time 25.0

Table 2. TCD, FID, Methanizer settings

TCD
Ar reference flow 10 mL/min
Temperature 200°C
Filament temperature 254°C
Carrier gas N2 /Ar
Total Time 25.0
FID
Temperature 300°C
Ar makeup flow 20 mL/min
H2 flow 10 mL/min
Air flow 300 mL/min
Methanizer
Temperature 400°C

Table 3. Valves

Time (min) (1) Gas Sampling Valve (2) Series bypass Sample Event A
Initial Fill Series OFF OFF
3.0 Fill Series OFF ON
4.2 Fill Bypass OFF ON

Results and Discussion

Chromatograms of both TCD and FID channels are shown in Figure 1 and Figure 2. The complete hardware configuration (SHS-40/TOGA) is shown schematically in Figure 3.

Repeatability is tested by analyzing multiple samples from the same source. Data can be found in Table 4.

Table 4. Repeatability Data

Run N2 Peak Area CH4 Peak Area CO2 Peak Area
1 692201 609 369764
2 696712 606 365757
3 669175 584 361535
4 678626 592 361783
5 709715 577 364403
6 702775 576 376105
7 724545 607 393602
n 7 7 7
Average 696249.9 593.0 370421.3
St.Dev. 18640 14.4 11414
RSD (%) 2.68 2.43 3.08

A graphic representation of the data is shown in Figure 4 and Figure 5.

Besides the analytical result also the window specified in the ASTM D 3612 method is presented. From the data presented in Table 4, Figure 4 and Figure 5 it is clear that the repeatability of the system is well within the window specified by the ASTM D 3612.

Figure 4. Repeatability results of a Transformer Oil.

Conclusion

Full separation of all components of interest, easy and reliable quantification results with very good repeatability was achieved.

The analysis of dissolved gases in transformer oil according to ASTM D 3612, Method C, can be performed perfectly with the Bruker Transformer Oil Gas Analyzer (TOGA Analyzer) in conjunction with the Bruker SHS-40 Automated Headspace Sampler.

About Bruker - Chemical and Applied Markets

Bruker is the new name in chemical analysis. Accurate and comprehensive analysis of exogenous and discrete elements in a wide range of sample matrices are key applications for many analytical chemistry groups. To address the needs and challenges of analysts working in those areas, Bruker has expanded their product family to provide, and expertly support, a series of fully integrated solutions including:

  • Gas Chromatography-Mass Spectrometers (GC/MS and GC/MS/MS)
  • Inductively Coupled Plasma Mass Spectrometers (ICP-MS)
  • Gas Chromatography Systems (GC)

Widely used in food and consumer safety testing, forensic, industrial, environmental, and clinical laboratories, these systems are well accepted and established market leaders that universally deliver outstanding performance at a premium value.

This information has been sourced, reviewed and adapted from materials provided by Bruker - Chemical and Applied Markets (CAM).

For more information on this source please visit Bruker - Chemical and Applied Markets.

Date Added: Jun 15, 2012 | Updated: Jun 11, 2013
Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Submit