Topics Covered
Introduction
Instrumentation Used
Materials and Reagents
Sample Preparation
Conditions
Results and Discussion
Conclusion
About Bruker - Chemical
and Applied Markets
Introduction
Insulating fluids, generally mineral oils, are used in transformers.
Under normal, mild conditions there is very little decomposition.
Occasionally however (localized or general) overheating of the oil
occurs and decomposition products are formed. If the concentration of
these gases reaches a critical point, the chances of catastrophic
transformer failure are high.
The ASTM D 3612 method describes in detail three different routes:
A. Vacuum Extraction
The gases are extracted from the oil via a vacuum extraction device
and analyzed via gas chromatography.
B. Stripper Column Extraction
Dissolved gases are extracted from a sample of oil by sparging the
oil with the carrier gas on a stripper column containing a high surface
area bead. The gases are then flushed from the stripper column into a
gas chromatograph for analysis.
C. Headspace Sampling
An oil sample is brought into contact with a gas phase (headspace)
in a closed vessel purged with Argon. As a result, a portion of a gas
dissolved in the oil is transferred to the Headspace.
This article describes analysis of a transformer oil by method C -
Headspace-Gas Chromatography
Instrumentation Used

|
| Figure 1.
TOGA Analysis, TCD Channel |
Materials and Reagents
“True North” DGA Oil Standard by Morgan Schaffer:
| Hydrogen |
88 ppm |
| Oxygen |
11163 ppm |
| Nitrogen |
40368 ppm |
| Methane |
96 ppm |
Carbon monoxide
|
89 ppm
|
Carbon dioxide
|
123 ppm
|
| Ethylene |
90 ppm |
| Ethane |
92 ppm |
| Acetylene |
84 ppm |
Sample Preparation

Figure 2. TOGA analysis FID
channel

Figure 3. Schematic
overview hardware.
The Morgan Schaffer Calibration Standard is carefully transferred
into
the headspace vial. The gases are extracted from the oil by means of a
headspace
sampler and injected onto a short Hayesep P precolumn and
then to a micro packed Carboxen-1000 column.
The fraction containing
Hydrogen, Oxygen, Nitrogen, Carbon Monoxide, and Methane will elute
direct from the Carboxen-1000 column to the micro packed Molsieve
column. Hydrogen, Oxygen and Nitrogen are detected by the TCD. Carbon
Monoxide, and Methane are detected by the FID, after passing the
Methanizer. When the Molsieve column is bypassed, Carbon Dioxide and
the C2-C3 isomers are eluting from the
Carboxen-1000 column and detected by the FID after passing the
Methanizer. The back flush time is set to completely elute the C3
isomers. C4 and higher are back flushed.
Conditions
Table 1. Column oven
settings
| Rate
(°C/min) |
Step
(°C) |
Time
(min.) |
| Initial |
50 |
5.0 |
| 10.0 |
75 |
0.0 |
| 20.0 |
220 |
10.25 |
|
Total Time |
25.0 |
Table 2. TCD, FID,
Methanizer settings
TCD
|
Ar reference flow |
10 mL/min |
| Temperature |
200°C |
| Filament temperature |
254°C |
| Carrier gas |
N2 /Ar |
| Total Time |
25.0 |
FID
|
Temperature |
300°C |
| Ar makeup flow |
20 mL/min |
| H2 flow |
10 mL/min |
| Air flow |
300 mL/min |
Methanizer
|
Temperature |
400°C |
Table 3. Valves
| Time
(min) |
(1) Gas
Sampling Valve |
(2)
Series bypass |
Sample |
Event A |
| Initial |
Fill |
Series |
OFF |
OFF |
| 3.0 |
Fill |
Series |
OFF |
ON |
| 4.2 |
Fill |
Bypass |
OFF |
ON |
Results and Discussion
Chromatograms of both TCD and FID channels are shown in Figure 1 and
Figure 2. The complete hardware configuration (SHS-40/TOGA) is shown
schematically in Figure 3.
Repeatability is tested by analyzing multiple samples from the same
source. Data can be found in Table 4.
Table 4. Repeatability Data
| Run |
N2
Peak Area |
CH4
Peak Area |
CO2
Peak Area |
| 1 |
692201 |
609 |
369764 |
| 2 |
696712 |
606 |
365757 |
| 3 |
669175 |
584 |
361535 |
| 4 |
678626 |
592 |
361783 |
| 5 |
709715 |
577 |
364403 |
| 6 |
702775 |
576 |
376105 |
| 7 |
724545 |
607 |
393602 |
| n |
7 |
7 |
7 |
| Average |
696249.9 |
593.0 |
370421.3 |
| St.Dev. |
18640 |
14.4 |
11414 |
| RSD (%) |
2.68 |
2.43 |
3.08 |
A graphic representation of the data is shown in Figure 4 and Figure
5.
Besides the analytical result also the window specified in the ASTM
D 3612 method is presented. From the data presented in Table 4, Figure
4 and Figure 5 it is clear that the repeatability of the system is well
within the window specified by the ASTM D 3612.

Figure 4. Repeatability
results of a Transformer Oil.
Conclusion
Full separation of all components of interest, easy and reliable
quantification results with very good repeatability was achieved.
The analysis of dissolved gases in transformer oil according to ASTM
D 3612, Method C, can be performed perfectly with the Bruker
Transformer Oil Gas Analyzer (TOGA
Analyzer) in conjunction with the
Bruker
SHS-40 Automated Headspace Sampler.
About Bruker -
Chemical and Applied Markets
Bruker
is the new name in chemical analysis. Accurate and comprehensive
analysis of exogenous and discrete elements in a wide range of sample
matrices are key applications for many analytical chemistry groups. To
address the needs and challenges of analysts working in those areas,
Bruker has expanded their product family to provide, and expertly
support, a series of fully integrated solutions including:
- Gas Chromatography-Mass Spectrometers (GC/MS and GC/MS/MS)
- Inductively Coupled Plasma Mass Spectrometers (ICP-MS)
- Gas Chromatography Systems (GC)
Widely used in food and consumer safety testing, forensic,
industrial, environmental, and clinical laboratories, these systems are
well accepted and established market leaders that universally deliver
outstanding performance at a premium value.

This information has been sourced, reviewed and adapted from
materials provided by Bruker - Chemical and Applied Markets (CAM).
For more information on this source please visit Bruker - Chemical and Applied Markets.