Site Sponsors
  • Handheld Thermo Scientific Niton XRF analyzers are engineered for portable elemental analysis
  • Goodfellow - Metals and Materials for Research and Industry
Site Sponsors

Buck 205 AA from Buck Scientific


The Buck 205 is the entry model AA from Buck Scientific.  Buck Scientific have removed the D2 Background correction and Three lamp turret to save money, the remainder of the instrument is identical to the 210VGP.  Many Customers with Low Dissolved solids in their samples use the 205 with great success.  The 205 is also popular amongst educators, since it is designed to be low maintenance and affordable. 

Specification

Electrical:


  • 110V AC  nominal (+10%), 50/60 Hz
  • 220, 240V AC, 50/60 Hz
  • Power Consumption: 50W

Optics:

  • Detector: model 928; wide range general purpose, 190-930nm
  • Optional Detectors: model 955; UV enhanced, wide-range, 190-930nm
  • model EMI9783B; narrow range furnace/hydride application, 165-600nm
  • Lenses:  Supracil - amorphous silica
  • Monochromator:  0.25m Ebert mount
  • Grating:  32nm x 27nm; 600 grooves/mm
  • Wavelength adjustment:  3 digit mechanical, 0 to 1000nm +1 nm
  • Reproducibility:  +0.2 nm
  • Resolution:  variable slit - 2Å, 7Å, and 20Å

Operating Modes:

  • Absorbance/Emission:  -0.0820 to 3.2000
  • Concentration:  to 5 significant digits
  • Integration Period:  0.5 to 10 seconds
  • Screen Refresh : 0.5 to 1.5 seconds
  • Recorder Output:  1V/ABS (-0.08 to 3.2V)
  • Giant Pulse (Self-reversal)

Hollow Cathode Lamps:

  • Dimension:  1.5" OD
  • Striking Voltage:  500V
  • Lamp Current:  0 to 18 mA average current
  • Duty Cycle:  25%
  • Modulation Frequency:  variable; 33 to 200 Hz (142 Hz Norm.)

Burner Assembly:

  • Design:  Polyethylene Pre-mix chamber,  glass impact bead dispersion
  • Burner Head:  Titanium; air-acetylene head - 4" x 0.026" single slot
  • (nitrous oxide head - 2" x 0.019" single slot)
  • Adjustments:  Horizontal g

Performance:

  • Average Noise (at 3σ):  0.0018 ABS (Cu at 324.7nm, 7Å slit, 5 sec. int.)
  • Sensitivity:  see specific element chart (Sect. 4)
  • Reproducibility:  <+5%

For more information, please contact Buck Scientific