Using Dry Heat for Sterilization

Dry heat is usually employed for sterilization instead of moist heat since certain materials are sensitive to moisture. The use of dry heat guarantees that glass and other lab equipment is free of pyrogenic materials, which are too small to be removed by filtration. Dry heat decomposes pyrogens into harmless atoms and molecules.

Dry Heat Sterilizer

Validation of a Dry Heat Sterilizer

In the case of a hot air tunnel or a chamber, the sterilization process is crucial and it is essential to perform regular validation. Validation is the documented procedure of acquiring, recording, and interpreting results to guarantee the consistent effectiveness of the dry heat sterilization process. A dry heat sterilizer is validated through accurate measurement of the temperature at crucial points inside the sterilization chamber.

In general, the dry heat process is carried out at a temperature of 250 °C to 400 °C for a differing amount of time. It is necessary for the sterilizer to heat all parts of its load to the predetermined temperature for a particular time period, which is adequately long to accomplish the desired sterility.

In order for all sensors to conform to international standards, they should be calibrated prior to and following the validation test. The produced data must be compiled to assess the potential of the sterilization tunnel to sterilize and depyrogenate vials of varying sizes.

AMETEK’s PTC-425 for calibrating sterilization tunnels could be advantageous to manufacturers in the medical, pharmaceutical, or food industry, as it has the perfect temperature range of 33 °C to 425 °C and an accuracy of ±0.13 °C.

With respect to heat tunnels, the PTC-425 calibrator serves three purposes:

  1. Calibration of heat tunnel validation sensors, both prior to and after the test
  2. Calibration of all fixed sensors in the tunnel
  3. Calibration of overheat security switch sensors

Moreover, PTC-425 is also designed with work order functionality, an exceptionally user-friendly navigation, high-speed heating and cooling times, and a large color display.

This information has been sourced, reviewed and adapted from materials provided by AMETEK Sensors, Test & Calibration.

For more information on this source, please visit AMETEK Sensors, Test & Calibration.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    AMETEK STC. (2019, October 18). Using Dry Heat for Sterilization. AZoM. Retrieved on November 28, 2022 from https://www.azom.com/article.aspx?ArticleID=17166.

  • MLA

    AMETEK STC. "Using Dry Heat for Sterilization". AZoM. 28 November 2022. <https://www.azom.com/article.aspx?ArticleID=17166>.

  • Chicago

    AMETEK STC. "Using Dry Heat for Sterilization". AZoM. https://www.azom.com/article.aspx?ArticleID=17166. (accessed November 28, 2022).

  • Harvard

    AMETEK STC. 2019. Using Dry Heat for Sterilization. AZoM, viewed 28 November 2022, https://www.azom.com/article.aspx?ArticleID=17166.

Tell Us What You Think

Do you have a review, update or anything you would like to add to this article?

Leave your feedback
Your comment type
Submit