Detailed hydrocarbon analysis (DHA) – Past, Present and Future Part 2

As an expert in the fuels analysis industry for over 30 years, Dan Wispinski of VUV Analytics has compiled a three-part series on the history of DHA, the current technique and future technologies.

Part two of this series focuses on the future of detailed hydrocarbon analysis. In the first article, Dan Wispinksi talked a little about his experience in shaping what DHA has become today.  That future uses vacuum ultraviolet spectroscopy (VUV) in an application called Verified Hydrocarbon Analysis™ (VHA) that maintains the advantages of DHA all while eliminating the drawbacks, resulting in more accurate data, increased productivity, and ease-of-use.

What is VUV Spectroscopy and VHA?

VUV Spectroscopy takes advantage of the VUV region of the electromagnetic spectrum which is characterized by high-energy and short wavelengths of light from 125nm to 240nm. Every compound absorbs here except for carrier gasses (i.e, hydrogen, helium, and argon). More importantly, every compound has a unique spectral fingerprint. These make it possible to accurately identify compounds of interest using a library of spectra. The VHA library covers most hydrocarbon compounds of interest.

One of the biggest challenges associated with the traditional 100-meter DHA method is achieving the separation necessary to ensure accurate identification of all compounds of interest – especially critical pairs (figure 1).  These are closely eluting compounds of very similar boiling points. To achieve separation, a tuning column is often required and run times can be over 140 minutes. Since VHA uses spectral fingerprints, critical pairs need not be completely separated; therefore, shorter columns can be used which accelerates analysis time while ensuring accuracy.

Misidentified or unidentified peaks using DHA occur because identification depends solely on retention time and operator interpretation -- many coelutions still occur.  Alternatively, VHA identifies compounds and classes not only by a retention window, but also by a compound’s unique spectral profile. The absorbance or response is directly proportional to concentration and allows accurate quantitation using similar normalization calculations as DHA.  VHA also uses the predictable response of analytes to eliminate calibration requirements. During VHA data processing, most gasoline compounds are “verified” by their unique spectra in the VHA library. When an exact match is not made, the hydrocarbon class and carbon number is assigned, made possible because compounds in a hydrocarbon class share spectral similarities. Every time interval is analyzed, resulting in zero “unknowns” compared to traditional DHA.

One of the core concepts of VUV Spectroscopy is that traditional peak integration is not performed. There are no peak starts, ends, tangent drops, or baseline considerations that DHA analysts spend significant resources on.  Even experienced operators may miss that the baseline follows valley to valley peaks points, omitting a large portion of FID response under a region of closely eluting compounds.  This is a common DHA error source.  Instead, VUV spectroscopy uses Time Interval Deconvolution™ (figure 2).  These time intervals, like time slices in GC distillation methods, are individually analyzed to solve the entire chromatogram. This eliminates peak integration and allows automated processing.

The future of hydrocarbon analysis is VHA.  This method provides the detailed analysis that is required while improving accuracy through spectral verification, increasing productivity from faster analysis time, less human intervention, and is easier-to-use.

Figure 1: Critical pairs deconvolved with VHA. Image Credit: VUV Analytics

Figure 2: Time Interval Deconvolution. Image Credit: VUV Analytics

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    VUV Analytics. (2020, June 03). Detailed hydrocarbon analysis (DHA) – Past, Present and Future Part 2. AZoM. Retrieved on September 20, 2020 from https://www.azom.com/article.aspx?ArticleID=19325.

  • MLA

    VUV Analytics. "Detailed hydrocarbon analysis (DHA) – Past, Present and Future Part 2". AZoM. 20 September 2020. <https://www.azom.com/article.aspx?ArticleID=19325>.

  • Chicago

    VUV Analytics. "Detailed hydrocarbon analysis (DHA) – Past, Present and Future Part 2". AZoM. https://www.azom.com/article.aspx?ArticleID=19325. (accessed September 20, 2020).

  • Harvard

    VUV Analytics. 2020. Detailed hydrocarbon analysis (DHA) – Past, Present and Future Part 2. AZoM, viewed 20 September 2020, https://www.azom.com/article.aspx?ArticleID=19325.

Tell Us What You Think

Do you have a review, update or anything you would like to add to this article?

Leave your feedback
Submit