Measuring Oxygen in Hazardous Areas — Process Oxygen Analyzers

The GPR-2800 range of process oxygen analyzers enables measurements from 0%–1% up to 0%–25% O2 in hazardous areas. In combination with Analytical Industries’ modular sample system, the gas analyzer will surely be appropriate for a majority of the installations for demanding oxygen analysis and will satisfy the requirements of process customers across the globe.

The proprietary long-life oxygen sensor technology from Analytical Industries enables a maintenance-free sensor operation of up to 32 months for low cost of ownership. The well-recognized and industry-certified electrochemical gas sensor quality guarantees the reliability of oxygen content measurements even in environments containing high carbon dioxide levels.

  • Four measurement ranges of 0%–1%, 0%–5%, 0%–10%, and 0%–25% O2
  • Sample in CO2 backgrounds with XLT sensor
  • Optional liquid drain option for the removal of entrained liquids
  • ATEX or UL certified for use in hazardous areas
  • Easy-to-use HMI
  • Low maintenance with an average sensor life of about 32 months (for GPR sensor)
  • Optional Modbus communications

Electrochemical Sensors from Analytical Industries have a long lifetime for minimum maintenance

Technical Specifications

GPR-2800 IS GPR-2800 AIS GPR-2800 AIS LD GPR-2800 IS LD
Measurement range 0-1%, 0-5%, 0-10% & 0-25% O2
Accuracy < 2% of selected range
Response time T90 < 10 seconds
Sensitivity (LDL) 0.05%
Linearity <1% of scale
Sensor model GPR-11-32 GPR-11-32-LD
Sensor life 32 months in air at 25 °C (77 °F) and 1 atm
Sensor model XLT-11-24 XLT-11-24-LD
Sensor life (in air 5% and 1 atm) 24 months in air
Calibration interval 30 days
Operating temperature GPR sensor: 5 °C to 45 °C (41 °F to 113 °F)
XLT sensor: -10° to 45 °C (14 °F to 113 °F)
Inlet pressure 0.34-2 barg (5-30 psig) with atmospheric vent
Flow rate 0.5-1.0 Nl/m (1-2 SCFH)
Gas connections 1/8" compression tube fittings 1/4" compression tube fittings
Wetted parts 316 Stainless steel
Display Graphical LCD 7 x 3.5 cm (2.75 x 1.375"); resolution 0.001
Enclosure NEMA Type 3R suitable for rain in outdoor applications (UL) NEMA 4x (ATEX)
Compensation Temperature
Signal output 4-20 mA 4-20 mA or 1-5V DC 4-20 mA
Communications Optional Modbus RTU (AIS - 24 V models only)
Alarms None Two user configurable alarms: magnetic coil relays rated 3A at 100 V AC None
Power 18-24 V DC (IS models)
12-28 V DC or 110-220 V AC (AIS models)
Area classification UL: Class I, Division 1, Groups C and D
ATEX: Ex II 2 G
Ex d [ib] ib IIB T4 Gb
Tamb - 20 °C to +50 °C"