The Relationship Between Oxidative Damage and a Scanning Electron Microscope Sample

O2 is a particularly reactive gas, and certain materials are incompatible with it. Oxidation processes are activated as quick as particular samples are exposed to the atmosphere. This impacts the characteristics and structure of the samples and in many cases can permanently cause damage.

This article outlines how oxidative damage can be avoided and how SEM analysis can be conducted on samples that are oxygen-sensitive without impacting the structure of the sample itself.

Samples and Oxidation

Oxidation reaction rates are highly contrasting, for example iron rusting takes years, whereas methane burns in a few milliseconds.

Lithium-based compounds are prone to being highly reactive when in contact with oxygen, which makes developing new materials for battery electrodes more complicated.

This can also be seen in Phosphor, Fluorine and many other elements of the periodic table. Certain compounds are so oxygen-sensitive that they can even start explosive reactions if ignition is present.

The growth of microorganisms is also strongly affected by oxygen deprivation. This is the parameter that separates the normal cellular breathing from the fermentation process or, sometimes, the inhibition of microbial growth.

The latter effect is mostly utilized in food preservation and production, where N₂, a gas with high chemical inertia, is employed to replace oxygen in packages of food, or to provide a safe environment at the production location in order to reduce the growth of pathogens.

As a side effect, this method also stops oxidation reactions in the food, keeping the product fresh and extending its shelf life.

How to Prevent Oxidative Damage to Samples in a Laboratory

A similar technique is frequently employed in laboratories, where the samples are prepared or manufactured in an environment free from oxygen. These are commonly referred to as ‘glove boxes’. They are large, sealed containers with a conditioned atmosphere where air is replaced by a gas with a high chemical inertia: typically Argon or N₂.

The decision between the two gasses typically depends on the nature of the sample and on the environment in which it is being prepared.

An example of a typical glovebox environment.

Image 1. An example of a typical glovebox environment.

When exposed to high tensions, N₂ can produce nitrogen radicals that can react with the sample and adjust its structure. Argon is a noble gas, which means its chemical inertia is higher than the N₂ gas, making it the ideal solution for glove box operation and conditioning.

As the cost of a glove box and the large amount of gas used to fill it can be considerable;the space inside the glove box is incredibly valuable. This is why desktop instruments are the only suitable choice when SEM analysis is needed.

On the contrary, Argon is very sensitive to high tension and is prone to creating sparks when high tension connections are exposed. This makes performing EDS analysis and imaging the samples a much more complicated task, as a high-voltage electron beam is incorporated.

Regardless of this, desktop SEMs that are not vulnerable to this effect are available and can deliver excellent images and correct EDS analysis of sensitive samples. Discover how by downloading the free Phenom ProX specification sheet.

This information has been sourced, reviewed and adapted from materials provided by Thermo Fisher Scientific Phenom-World BV.

For more information on this source, please visit Thermo Fisher Scientific Phenom-World BV.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Thermo Fisher Scientific Phenom-World BV. (2023, March 11). The Relationship Between Oxidative Damage and a Scanning Electron Microscope Sample. AZoM. Retrieved on May 05, 2024 from https://www.azom.com/article.aspx?ArticleID=18281.

  • MLA

    Thermo Fisher Scientific Phenom-World BV. "The Relationship Between Oxidative Damage and a Scanning Electron Microscope Sample". AZoM. 05 May 2024. <https://www.azom.com/article.aspx?ArticleID=18281>.

  • Chicago

    Thermo Fisher Scientific Phenom-World BV. "The Relationship Between Oxidative Damage and a Scanning Electron Microscope Sample". AZoM. https://www.azom.com/article.aspx?ArticleID=18281. (accessed May 05, 2024).

  • Harvard

    Thermo Fisher Scientific Phenom-World BV. 2023. The Relationship Between Oxidative Damage and a Scanning Electron Microscope Sample. AZoM, viewed 05 May 2024, https://www.azom.com/article.aspx?ArticleID=18281.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.