.jpg)
Topics Covered
Introduction
Advantages of Using NMR to Test
Fat Content in Foods
Standard Wet Chemistry Methods Result
in Bottlenecks
NMR and Its Advantages over Other
Techniques
NMR Selected to Solve Throughput
Challenges
Verifying if the Oxford Instruments MQC Benchtop
NMR is up to the Task
Throughput Results
Achieved
Replacing Soxhlet Units with MQC Benchtop NMR
Instruments
Calibrating the Oxford
Instrument MQC Benchtop NMR Instruments
Getting the NMR Test Method Accredited
NMR Eliminating Quality Control Bottlenecks
Introduction
The quickening pace of business caused one of Europe's leading independent
testing laboratories to seek an alternative to the standard solvent
extraction/acid hydrolysis (Soxhlet) method for testing the amount of fat in a
variety of foods. This contract laboratory, with numerous sites throughout the
UK and Ireland, provides quality control analytical services to the food
industry. By converting from the wet chemistry method to an MQC benchtop
Nuclear Magnetic Resonance (NMR) analyzer for measuring fat content of
foods, the lab reaped significant economic and environmental benefits. The box
on the right is a brief overview of NMR's
advantages over other secondary methods.
Advantages of Using NMR to Test Fat Content in Foods
Key features of NMR include:
- Can be calibrated to cover a concentration range from 0.5 to 100 percent fat
- Primary calibration can be produced using a single fat sample
- Requires infrequent recalibration
- Sample measurement time is short
- Minimal sample preparation necessary
- No solvents are required
- Excellent for bulk measurements
- Insensitive to sample granularity and
- product additives
- Non-destructive, facilitating repeatability measurements
Standard Wet Chemistry Methods Result in Bottlenecks
Customers send foodstuff samples to this contract quality control laboratory,
which has a specialty in fast turnaround service. A typical request includes
five or six measurements, including fat (oil) content. The Soxhlet method used
for the oil measurement is slow, with measurements taking as long as 6 hours.
This situation led to serious bottlenecks that were reducing throughput and
affecting the lab's ability to deliver its promised rapid analysis service. The
process is also rather cumbersome, can be inaccurate, and requires highly
skilled personnel. In addition, many of the hazardous chemicals used are
becoming increasingly unacceptable according to international environmental
standards.
NMR and Its Advantages over Other Techniques
The lab began seeking a rapid technique that would improve turnaround time
without increasing operating costs, but would also be comparable to the industry
standard Soxhlet technique.
There are a number of analytical methods that can be used to conduct the
testing. Such methods are often referred to as secondary techniques, since they
are usually set-up to match the results produced by solvent extraction. To
provide a result equivalent to the traditional extraction techniques, secondary
techniques must be correlated against the reference technique used. Although
they may be fast in use, many secondary techniques need to be calibrated and
maintained regularly. Maintenance and consumables add significantly to the cost
of ownership.
For example, Supercritical Fluid Extraction (SFE) is reasonably fast, but it
requires high maintenance and the cost of compressed CO2 that is used
to extract oil is also significant. Near Infra-Red (NIR) is sometimes used, but
it is generally sensitive to the surface rather than the bulk of the sample, and
has substantial calibration and calibration maintenance issues. NIR calibration
is complex because measurements are sensitive to product granularity and other
physical characteristics and can be affected by additives such as seasoning,
making it difficult to maintain accurate calibrations on a large variety of
product types. This gives NIR limited applicability for the quality control of
fat content in foodstuffs.
In contrast to the standard wet chemistry methods and various secondary
techniques, low
field nuclear magnetic resonance (NMR) provides a fast, direct and user
friendly method for determination of the fat and oil content in foodstuffs. The
technique is based on measurement of the NMR response
obtained from fat in the product, and quantification of the fat content by
simple and direct calibration without the use of chemometrics. The instrument is
extremely easy to operate and does not require the use of skilled chemists or NMR
specialists.
NMR can
be calibrated to cover a concentration range from 0.5 to 100 percent fat. The
user can produce a primary calibration using a single sample of fat. NMR is very
stable over the long-term, and therefore requires infrequent recalibration.
Sample measurement time is short, typically about 20 seconds, allowing a high
throughput of samples and efficient laboratory operation. Minimal sample
preparation is required because the entire sample is normally loaded into a tube
and measured directly, and there are several different size tubes available.
With NMR, no solvents are required since the sample is analyzed in
its natural state. The instrument facilitates bulk measurement; the signal is
generated from the whole sample, ensuring that the result embodies everything
inside the sample, not just on the surface. NMR is
virtually insensitive to sample granularity and such additives as spices,
flavors, colours and salt. Finally, unlike Soxhlet, the NMR technique
is non-destructive, so any required repeatability measurements can be made
easily.
NMR Selected to Solve Throughput Challenges
After reviewing the potential alternative solutions to bottlenecks associated
with fat measurements, the laboratory contacted Oxford
Instruments, which offers a benchtop NMR
instrument widely used in industry to measure the oil content in foodstuffs
and oil seeds. Oxford Instruments recommended its powerful but compact MQC benchtop NMR
analyzer for this application because it offers the analyst the benefits of
accurate quantitative results, combined with sampling ease and convenience.
Verifying if the Oxford Instruments MQC Benchtop NMR is up to
the Task
To verify whether the MQC NMR instrument would meet the laboratory's needs,
applications specialists from Oxford Instruments collected and tested samples of some of the
foods the lab typically analyzes and compared NMR
measurements on dried samples with fat measurement values obtained with the
Soxhlet method also after oven drying. The goals of the testing process were to
analyze 80 percent of samples using the MQC NMR
instrument, achieve correlation to within 5 percent of the wet chemistry
method, and achieve a repeatability of within 5 percent.
Applications specialists sampled a range of foods, with fat contents ranging
from 2.1 to 40.2 percent by mass, including baked cheese, muesli, milk powder,
chicken powder, trifle, garlic bread, macaroni and cheese, meat, and chicken
sandwich filler. Sampling was conducted using an Oxford Instruments
MQC-23 benchtop NMR analyzer equipped with a 26 mm diameter probe.
As shown on Table 1, the NMR results compared very closely to wet chemistry.
Table 1. Comparison of wet chemistry and NMR-MQC
instrument
| Sample |
Given fat content % (wet
chemistry) |
Measured fat content % (NMR – MQC
instrument) |
| Baked cheese |
6.1 |
5.8 |
| Muesli |
2.3 |
2.2 |
| Milk Powder |
25 |
24.2 |
| Chicken powder |
40.2 |
40.1 |
| Garlic Bread |
16.1 |
16.0 |
| Macaroni cheese |
2.8-3.4 |
3.4 |
| Meat |
9.9 |
9.2 |
| Sandwich filler |
21.7 |
21.6 |
Figure 1 shows a calibration for the samples, indicating an excellent linear
correlation between the NMR response and the concentration of fats in the products.
Instrument repeatability was tested by measuring one sample ten times. Each
sample was conditioned by placing it in a controlled-temperature heating block
at 40 °C for 20 minutes prior to measurement. NMR is
temperature sensitive, and with a stabilised magnet temperature of 40 °C,
repeatability and precision are optimised by pre-conditioning the sample at that
temperature.
In addition, some samples must be heated to melt the fat so it becomes
visible to the NMR.
.jpg)
Figure 1. Fat (Oil) Content in Foodstuffs. Calibration
obtained for fat content in foodstuffs; standard deviation of the linear fit is
0.20 % by mass, correlation coefficient R2 = 1.00. Measurements were made using
an Oxford Instruments MQC-23 benchtop NMR analyzer equipped with a 26 mm
diameter probe.
Table 2 illustrates the repeatability test results, which were well within
the anticipated range.
Table 2. Results of instrument repeatability test
|
Given fat content % (wet
chemistry) |
Results (%) of Repeat NMR
Measurements |
Mean Value % |
Standard Deviation
% |
|
16.1 |
15.79 |
15.80 |
15.80 |
15.81 |
15.82 |
15.81 |
15.82 |
15.82 |
15.83 |
15.81 |
15.81 |
0.01 |
Throughput Results Achieved
Laboratory managers reported that the testing results exceeded their
expectations, and the group purchased an MQC low-field,
benchtop NMR analyzer. Currently, about 90-95 percent of the laboratory's
samples are run on the MQC instrument and the lab has increased its efficiency
because of the rapid measurement capability.
Table 3 compares the number of analyses per hour and the skills required for
the NMR technique versus that of Soxhlet. According to Kulsum Jassat, sales
manager for the UK and Europe, "Throughput has increased from 80 to 200 samples
per day. With solvent extraction, each sample took approximately six hours from
set-up to completion, or a rate of about one sample per hour. With NMR, it is
usually less than one minute per sample after sample set-up." She adds, "The
company achieved a return on investment (ROI) within 5 months of purchasing the
instrument."
Table 3. Measurement comparison
|
Technique |
Analyses per hour at room
temperature |
Skills
required |
|
Solvent extraction |
1 |
Highly skilled operator, usually a chemist |
|
MQC/NMR |
60 (1 per minute) |
Non-skilled operator |
Replacing Soxhlet Units with MQC Benchtop NMR
Instruments
The purchase of the NMR MQC instrument allowed the laboratory to replace four of
its Soxhlet units. This significantly reduced the cost of purchase and disposal
of solvents, increased available laboratory space, reduced running costs,
achieved environmental benefits associated with reduced solvent usage, and
facilitated better deployment of laboratory staff.
Calibrating the Oxford Instrument MQC Benchtop NMR
Instruments
Since there is a slight calibration drift over about six months to a year,
the laboratory will still need to recalibrate periodically against the primary
technique or with set-up samples supplied by Oxford Instruments. During
installation, set-up samples can be made to physically capture and maintain the
original calibration, saving time during instrument recalibration. They also
save time by allowing calibrations to be transferred to other similarly
configured Oxford MQC NMR instruments.
Getting the NMR Test Method Accredited
While not required by most potential users of NMR
instruments, this particular contract lab chose to get its new fat measurement
method accredited by the United Kingdom Accreditation Service (UKAS); the
national accreditation body that uses a set of internationally agreed standards
to assess organizations that provide certification, testing, inspection, and
calibration services.
The lab developed a universal method using a sample blending process, sample
conditioning to ensure the optimal temperature for reading the NMR signal,
followed by analysis using the MQC
instrument. They used hundreds of data samples from Soxhlet and NMR-MQC to
prove that the results obtained from the MQC
reproducibly fell within acceptable tolerance. This method has been
UKAS-validated and fully accredited since 2009.
NMR Eliminating Quality Control Bottlenecks
NMR
provides a fast and reliable method to increase throughput for quality control
testing of fat content in foodstuffs. Although past history may have placed the
venerable Soxhlet solvent extraction technique in the role of industry standard,
for those seeking to eliminate quality control bottlenecks, the MQC NMR
analyzer is poised to take over the lead.
Source: Oxford Instruments Magnetic Resonance
For more information on this source, please visit Oxford Instruments
Magnetic Resonance