Using XRF to Analyze Total Chlorine at 1500 W

A calibration curve has been created in accordance with ASTM D4929 method to show the analysis of chlorine (Cl) in oil. The only difference in method was the use of a rhodium anode tube.

Instrumentation

The powerful and stable Thermo Scientific™ ARL™ PERFORM’X WD-XRF instrument has been employed. An alternative for rapid change-over of the analysis environment from vacuum to helium atmosphere (< 6 min.) is present which enables liquids analysis.

Below the analytical conditions are displayed.

The Analysis of Chlorine in Oils

The ASTM D4929 method has been selected for low Cl in oil. For a calibration curve to be created, five standard samples were arranged in accordance with the standard.  This method proposes using net peak intensities, therefore a background position was selected.

The background position and the counting time on the Cl peak are the same. From a concentrate product which contained 1000 ppm of Chlorine a mother solution was prepared by diluting this concentrate with Iso-Octane.. Following this, the mother solution was utilized to formulate multiple samples with differing Chlorine concentrations.

Table 1. The Instrument Settings. Source: Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers

Instrument ARL PERFORM’X
Crystal Ge 111
Collimator 0.40
Detector FPC
kV – mA 30 – 50
Power 1500 watts
Time / Line 1 180 sec
Total time 360 sec

 

Each standard net intensity (peak minus background) is linked with the Cl content in ppm, resulting in the calibration curve shown in Figure 1. Table 2 displays the numerical results containing the absolute disparity between nominal and determined concentrations, and the Standard Error of Estimate.

Table 2. Calibration results for five standard samples with low Cl content. Source: Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers

Sample # Cl intensity Nom. conc. ppm Calc. conc. ppm Absol. diff. ppm
0 108 0 0.3 0.3
1 167 7.9 8 0.1
2 239 17.7 17.5 0.2
3 409 40.4 39.7 0.7
4 572 60.4 61 0.6
5 655 71.9 71.8 0.1

SEE = 0.51 ppm
Correlation error R2 = 0.999753

Cl in Oil - Low Levels

Cl calibration curve.

Figure 1. Cl calibration curve. Image Credit: Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers

Repeatability Test on Oil Samples Containing Cl

One oil sample with 1.4 ppm level has been utilized for a repeatability test for the analysis of Cl. Six liquid cells were arranged and examined. The results found are listed in the table below.

Table 3. Repeatability test. Source: Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers

Sample Cl concentration in ppm
Cell 1 1.2
Cell 2 1.5
Cell 3 1.2
Cell 4 1.5
Cell 5 1.2
Cell 6 1.8
Average 1.4
SD 0.6

 

Conclusion

It is possible to obtain good calibration curves using the ARL PERFORM’X XRF for Cl determination in oil when employing the appropriate ASTM standard method despite the use of a rhodium anode tube. Due to the ARL Perform’X innovative helium shutter shielding the goniometer chamber from the helium environment and the reproducibility of loading, superb repeatability of analysis can be exhibited for Cl analysis at low level.

The results found demonstrate that excellent accuracy and precision is achieved with the ARL PERFORM’X Sequential XRF instrument. This instrument works well for the assessment of Cl in petrochemical products.

Thermo Scientific ARL PERFORM’X Sequential XRF.

Thermo Scientific ARL PERFORM’X Sequential XRF. Image Credit: Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers

This information has been sourced, reviewed and adapted from materials provided by Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers.

For more information on this source, please visit Thermo Fisher Scientific - Elemental Analyzers and Phase Analyzers.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Thermo Fisher Scientific - Elemental and Phase Analysis. (2020, November 09). Using XRF to Analyze Total Chlorine at 1500 W. AZoM. Retrieved on April 26, 2024 from https://www.azom.com/article.aspx?ArticleID=19805.

  • MLA

    Thermo Fisher Scientific - Elemental and Phase Analysis. "Using XRF to Analyze Total Chlorine at 1500 W". AZoM. 26 April 2024. <https://www.azom.com/article.aspx?ArticleID=19805>.

  • Chicago

    Thermo Fisher Scientific - Elemental and Phase Analysis. "Using XRF to Analyze Total Chlorine at 1500 W". AZoM. https://www.azom.com/article.aspx?ArticleID=19805. (accessed April 26, 2024).

  • Harvard

    Thermo Fisher Scientific - Elemental and Phase Analysis. 2020. Using XRF to Analyze Total Chlorine at 1500 W. AZoM, viewed 26 April 2024, https://www.azom.com/article.aspx?ArticleID=19805.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.