Posted in | News | Business

Improved Humidity Conformance Testing for DVS Analyzers

Hiden Isochema announces a new improved conformance test method for their dynamic vapor sorption (DVS) analyzers.

 

 

The traceable method verifies the combined temperature and relative humidity measurement and control for all gravimetric analyzers with a humidifier or vapor generator, including the IGAsorp, and relevant options of their IGA and XEMIS series.

The test involves monitoring the gravimetric response of a certified standard or saturated salt solution as the applied humidity is controlled near the equilibrium relative humidity (ERH). The Hertz-Knudsen equation is then used to determine the ERH of the standard and the combined measurement uncertainty. Conformance test certificates are automatically produced from HIsorp software using the calculated pass/fail criterion.

The method also includes options to cycle the test for verification of long term analyzer performance and also to change the test temperature if appropriate. The new improved method includes an automated pre-conditioning stage and may equally be applied to humidity feedback control and constant flow modes of dynamic vapor sorption analysis.

A low cost user test kit is available that also adds a weight conformance check for complete instrument verification in compliance with site quality systems including GMP.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Hiden Isochema. (2023, April 14). Improved Humidity Conformance Testing for DVS Analyzers. AZoM. Retrieved on April 26, 2024 from https://www.azom.com/news.aspx?newsID=45198.

  • MLA

    Hiden Isochema. "Improved Humidity Conformance Testing for DVS Analyzers". AZoM. 26 April 2024. <https://www.azom.com/news.aspx?newsID=45198>.

  • Chicago

    Hiden Isochema. "Improved Humidity Conformance Testing for DVS Analyzers". AZoM. https://www.azom.com/news.aspx?newsID=45198. (accessed April 26, 2024).

  • Harvard

    Hiden Isochema. 2023. Improved Humidity Conformance Testing for DVS Analyzers. AZoM, viewed 26 April 2024, https://www.azom.com/news.aspx?newsID=45198.

Tell Us What You Think

Do you have a review, update or anything you would like to add to this news story?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.