Multiple ASTM methods address flash point measurement for petroleum materials and related products. Among these, ASTM D93 Procedure A is used to specify diesel and aviation turbine fuel.

Image Credit: Marykor/Shutterstock.com
In this procedure, 70 mL of a sample is heated in a closed cup at a predetermined temperature rate. To determine the so-called ‘closed cup flash point’, a test flame is lowered into the vapor space at regular intervals.
Refinery laboratories face strict regulations that forbid the use of open flames in the vicinity of flammable liquids. In light of this, instrument manufacturers have faced the challenge of creating a safer flashpoint method.
The new method they created uses a small, continuously closed cup, an electric arc, and a smaller sample, in turn reducing waste and method turnaround time.
Now, a portable, rugged, and automated apparatus is available: ASTM D7094. Even for contaminated samples, there is no bias between the new ASTM D7094 and D93A.
Background
ASTM D93, commonly known as the Pensky-Martens method, is so named for the two German chemists who created both the procedure and its apparatus during the nineteenth century.
Under this conventional closed-cup approach, a 75 mL specimen is heated at a controlled rate. At set intervals, either a defined test flame or a heated wire is introduced into the vapor space of the momentarily opened cup. The temperature at which ignition occurs is then recorded as the “closed cup flash point.”
Because of the difficulties in ensuring safety, developing a safer flash point technique was a significant challenge for producers of petroleum testing instruments. In 1999, a new procedure was issued as ASTM D6450-99 to address this challenge.
This newer technique requires only a 1 mL specimen within a continuously sealed cup. Rather than relying on an open flame, it initiates ignition with an electric arc and pumps air into the sealed chamber. Flash point is identified through a pressure rise inside that chamber, while the heating rate matches that of D93A.
ASTM D6450 is also referred to as the continuously closed cup flash point (CCCFP) method. Its advantages include reduced waste from the smaller specimen volume, simpler cleaning, substantially shorter analysis time, and the possibility of producing a portable, durable, and fully automated instrument.
An interlaboratory investigation was carried out in 1997 under the designation RR:D02-1464. Its specimen group included three pure chemicals, one pure hydrocarbon, three fuels, and three lubricating oils. Each material was evaluated using D93A and the proposed technique that subsequently became D6450.
The study found reasonable correspondence between ASTM D93A and the proposed procedure. However, the repeatability and reproducibility observed for D93A were notably poorer than their published values. Using the study results, precision values for the new CCCFP technique were established, and the procedure was subsequently approved and issued as ASTM D6450-99.
Later experience revealed that specimens contaminated with small quantities of substantially lower-flash-point substances, such as diesel containing traces of gasoline or lubricating oil containing diesel fuel, produced D6450 results above those obtained with D93A. Depending on the contaminant type and concentration, these discrepancies were as high as 8 °C.
In 2000, the manufacturer of the D6450 instrument notified ASTM of this offset. The company described investigations aimed at adjusting the D6450 procedure to lessen the divergence from D93A, even for contaminated materials. Consequently, the CCCFP Task Group remained active.
In December 2000, findings were presented indicating that modifying selected D6450 parameters could eliminate the discrepancy between D93 results and those for contaminated specimens.
Following a preliminary investigation, the ASTM Flash Point Task Group agreed to organize an interlaboratory study. The resulting data would determine whether D6450 required revision. The group further agreed that the specimen group should contain contaminated fuels and lubricating oils, and that testing would use D6450-99, D93A, and the revised CCCFP procedure later issued as ASTM D7094.
Sample Set Design
The study included 13 specimens:
- Jet A (1)
- Jet A (2)
- Diesel (1)
- Diesel (2)
- Diesel (3)
- Lubricating oil (1)
- Lubricating oil (2)
- Two-stroke oil
- Anisole
- Jet A (1) with 0.25 V% gasoline
- Jet A (2) with 0.4 V% gasoline
- Jet A (2) with 2.0 V% biodiesel
- Diesel (2) with 0.3 V% gasoline
- Diesel (1) with 0.5 V% gasoline
- Lube oil (1) with 3 V% diesel
Contaminated materials were produced by volumetrically adding the specified proportion of the contaminant to the neat material stored in large drums. The containers were rolled to ensure homogeneity, after which the specimens were transferred into 250 mL glass bottles.
Each of the nine participating laboratories received three 250 mL bottles per specimen, one allocated to each procedure, so that every method used a fresh specimen.
Precision
Richard M. Stanley of Chevron Information Tech Co. calculated the precision results.
For the ASTM D93A Pensky-Martens procedure, the result distributions for both lubricating oils and the Jet A specimen containing biodiesel were clearly bimodal. These materials were therefore treated as outliers and excluded.
The precision estimates for the remaining specimens were:
- Repeatability, r = 4.1 °C
- Reproducibility, R = 6.9 °C
Relative to the published reproducibility of 0.078 times the flash point in °C, the D93A precision obtained in this program was greater for every specimen except one.
For the revised CCCFP procedure, ASTM D7094, the estimates were:
- Repeatability, r = 4.1 °C
- Reproducibility, R = 5.5 °C
When the two lubricating oil specimens were excluded, as had been done for the D93A precision calculation, precision improved to r = 3.8 °C and R = 5.4 °C.
Comparison to ASTM D93A
Richard M. Stanley of Chevron Information Tech Co. evaluated the flash point procedures used in this program according to ASTM D6708, 'Standard Practice for Statistical Assessment and Improvement of the Expected Agreement Between Two Methods that Purport to Measure the Same Property of a Material.'
The comparison range was limited to approximately 40–145 °C because the two higher-flash-point specimens (the two lubricating oils) were excluded as outliers from the ASTM D93A dataset.
ASTM D7094 and ASTM D93A Comparison
A limited, random, specimen-specific bias was observed. According to the ASTM D6708 analysis, no bias correction was necessary.
For the established ASTM D93 procedure, the repeatability and reproducibility determined in this program were considerably poorer than the published values. This outcome is explained, at least in part, by the substantial proportion of contaminated specimens included in the study.
Materials with relatively high flash points that contain small amounts of much lower-flash-point substances are recognized as especially challenging in flash point testing.
For D93A, however, the two specimens with flash points of 200 °C or higher were removed from the analysis because their results were bimodally distributed. Since these were neat lubricating oils, contamination cannot explain that behavior.
Summary
Results from the interlaboratory study indicate that for neat fuels, oils, and chemicals, ASTM D6450-99 produces values closely aligned with those from D93A. The revised CCCFP procedure, ASTM D7094, resembles D6450 but employs different operating conditions; like D6450, it agrees closely with D93A for neat fuels, oils, and chemicals. Importantly, D7094 also shows no bias relative to D93A for contaminated specimens.
In fact, the ASTM D6708 analysis demonstrated that, for the specimen group used in the interlaboratory study, no statistically meaningful bias exists between ASTM D7094 and D 93A.
For laboratories looking to take advantage of these modern flash point methods, the Grabner FPV Pro is the go-to solution for testing according to ASTM D6450 and ASTM D7094. By bringing these methods together in a modern automated platform, the FPV Pro enables laboratories to benefit from the inherently small sample requirements, continuously closed-cup design, reduced waste, simpler cleaning, and faster analysis associated with CCCFP testing, making it an ideal choice for efficient routine flash point measurement of fuels, oils, chemicals, and potentially contaminated samples.

This information has been sourced, reviewed, and adapted from materials provided by Grabner Instruments.
For more information on this source, please visit Grabner Instruments.