Infrared spectroscopy has played a major role in fuel analysis since the earliest stages of fuel characterization. Gasoline, diesel, and jet fuel are complex hydrocarbon mixtures, and many of their components do not produce distinctive infrared absorption. These components form the “matrix” or “background” of the spectrum.

Image Credit: Nattapol_Sritongcom/Shutterstock.com
By contrast, aromatics, olefins, and the numerous additives incorporated into fuels produce measurable absorption bands that can be used for characterization. Regulatory authorities closely monitor aromatic compounds such as benzene, and standardized infrared procedures are available to determine their levels, including ASTM D6277 and ISO EN238.
Additives used to improve fuel performance are also subject to control, both to support optimized blending and to address environmental concerns. Numerous compounds have been developed to raise oxygen content and promote more complete combustion. Because oxygen–carbon bonds interact readily with infrared radiation, IR spectroscopy became a preferred technique for identifying such materials, for example under ASTM D5845.
At present, the US Environmental Protection Agency (EPA) lists nearly 7400 substances supplied by different manufacturers and vendors for use as fuel additives.
Typically, a defined set of additives is selected to adjust key fuel characteristics, such as octane number in gasoline or cetane number in diesel. Over time, it became clear that a fuel’s spectrum should correlate with the additives that govern its performance. Because conventional standardized procedures for determining fuel properties, such as CFR knock engines, can be slow and costly, chemometric approaches that connect infrared data with fuel properties gained popularity.
During the 1990s, Dr. Werner Grabner, founder of Grabner Instruments in Vienna, Austria, introduced the IROX, the first portable FTIR spectrometer capable of determining oxygenate concentrations in gasoline using mid-infrared radiation. A specialized diesel instrument followed in 2000.
Both instruments determine the concentrations of multiple fuel constituents and estimate conventional fuel properties through prediction models developed for different geographic markets.
By the middle of the last decade, established benchtop infrared manufacturers recognized fuel testing as a new commercial opportunity and created comparable dedicated applications for fuel analysis. They also adopted Raman spectroscopy as a complement to conventional IR and introduced attenuated total reflection (ATR) infrared spectroscopy.
FTIR, Raman, and ATR are fundamentally different approaches to examine the same spectral information from somewhat different perspectives. No technique is universally superior; each offers advantages and limitations depending on its intended use. ATR is well-suited to online process applications, while FTIR instruments are generally compact, rugged, and appropriate for mobile fuel laboratories.
The decisive differences arise in terms of usability, technical support, flexibility, breadth of applications, design quality, durability, and the precision and accuracy of the instrument. Accuracy and precision depend directly on spectrometer design, resolution, stability, spectral range, and the intensity of the infrared source.
Concentration Measurement and Property Prediction
Using modern IR instruments designed for fuel analysis, a single measurement can provide the concentrations of several fuel components and properties such as octane number, distillation characteristics, or vapor pressure within minutes.
This is far faster than the potentially costly and technically demanding procedures specified in fuel standards. Infrared determination of substances such as additives is covered by standards including ASTM D5845, D6277, EN 238, and EN 14078.
Establishing correlations for broader fuel properties, however, is much harder to standardize. Fuel is a highly complex mixture that varies substantially among countries, making a universal, one-size-fits-all prediction model impractical. This raises several important questions.
Is Property Prediction a Measurement in the Scientific Sense?
In short, no. Spectroscopy employs electromagnetic radiation to induce electronic transitions in matter, often enabling identification of a substance and determination of its concentration within a mixture. An octane number, however, is not a concentration.
How Precise Is Property Prediction?
Precision depends on the quality of the acquired spectra and on how accurately the reference properties were measured. For this reason, numerous fuel spectra are collected, evaluated, classified, and stored in a database together with their known properties.
Principal component analysis (PCA) identifies spectral regions that correlate most strongly with a given property, enabling development of a prediction model. When an unknown fuel is measured, the IR spectrometer can apply that model to estimate RON, distillation characteristics, and other properties.
Can Property Prediction Match the Accuracy of Standardized Methods?
Yes and no. If the reproducibility (R) of a property measured by the conventional standard is no better than ±1, the prediction cannot achieve an R better than ±1. However, repeatability (r), defined as the standard deviation obtained when the same specimen is measured repeatedly on the same instrument, can exceed that of the original standard. Prediction precision also depends on the model generated from the database.
The more strongly the selected spectral regions correlate with a property, the better the prediction. Robustness likewise depends on the chemometric techniques used. Today, cluster analysis, multiple linear regression, and partial least squares methods are often applied together to extract the greatest value from the available data.

Figure 1. Motor octane number: predicted vs. CFR. Image Credit: Grabner Instruments
How Can a Prediction Achieve Robustness?
Over the past 15+ years, near-infrared (NIR) instruments have entered the market because they can support robust predictions and permit simpler instrument designs. These devices are often compact and portable and may employ longer absorption paths.
In this spectral region, fuel spectra consist largely of overtone vibrations, particularly C–H, O–H, and N–H bonds, and combinations of rotational and vibrational modes. Distinct absorption bands are therefore difficult to resolve but robust correlations with fuel properties can still be developed from the overall spectral pattern.
Because NIR spectra are less information-rich, such instruments generally do not quantify individual substances, and developing a prediction model requires a larger set of diverse fuels.
The mid-infrared region is often called the fingerprint region of the IR spectra. Individual substances generally display well-separated bands there, allowing identification based on band positions and concentration determination.
Filter-based instruments are often used when only a limited number of compounds, or a single target such as ethanol in gasoline or biodiesel (FAME) in diesel, are to be measured. Their use cases are limited, but they are usually inexpensive, compact, simple to operate, and capable of developing prediction models from correlations between selected spectral ranges and common fuel properties.
As with NIR systems, the restricted number of spectral ranges, corresponding to the number of filters, requires more collected data to build a dependable prediction model.
A limitation inherent to filter-based IR instruments is their inability to measure the baseline of a spectral band. When determining concentration, the background contributed by the fuel matrix and the instrument’s optical components can affect accuracy. Consequently, IR spectrometers of this design use dual-beam systems to measure incoming (Iin) and transmitted (Iout) infrared radiation simultaneously.
Physics Can’t Be Fooled
Certain physical constraints determine the usable spectral range of infrared spectrometers. The optical components and the heat generated by the IR source are fundamental to interferometer design and restrict both wavelength range and achievable resolution.

Figure 2. Interferogram retardation and maximum resolution. Image Credit: Grabner Instruments
First, the beam splitter material and cell windows must transmit infrared light without absorbing it. This becomes increasingly challenging as the wavelength range expands.
Most conventional mid-IR FTIR spectrometers use a similar design. Zinc selenide (ZnSe), a salt material, is commonly employed as the beam splitter. It transmits IR radiation from 2.5 μm (4000 cm-1) to 15.5 μm (645 cm-1), functioning essentially as a broad-pass filter for infrared waves.
Extending the wavelength range requires either widening this “filter” or changing it according to the wavelength being measured. The latter approach is expensive, cumbersome, and impractical for routine testing. A better solution is a specialized material that resists degradation while transmitting IR radiation from 1.25 μm (8000 cm-1) to 25 μm (400 cm-1).
Second, spectral resolution is related to the maximum retardation (δ, in cm) of the interferogram. Beam intensity is greatest when both mirrors are equidistant from the beam splitter and the two portions of IR light have no phase difference. This is known as zero retardation or zero path difference (ZPD).
For a given wavelength, intensity reaches a minimum when the moving mirror has traveled a distance (δ) equal to half the wavelength (v); at that point, the maximum of one wave aligns with the minimum of the other, causing cancellation. Because IR sources generate polychromatic radiation across a broad wavelength distribution, the interferogram peaks at δ=0, after which its envelope declines rapidly.
The distance δ, in cm, at which the envelope falls to the signal-to-noise limit determines the best resolution through its inverse, δ (1/δ), in cm-1. If interferogram data extends to 0.5 cm, the resolution is 2 cm-1.

Figure 3. Spectral range of IR sources according to Planck´s Law. Image Credit: Grabner Instruments
Third, according to Max Planck’s laws governing thermal radiation, the wavelength distribution emitted by an IR source broadens as its temperature increases. A comparable effect occurs when a light bulb is dimmed: at full voltage the light appears white and bright, while reducing the voltage shifts its color toward red, the lower-energy end of the spectrum.
Glowing Kanthal wires have commonly been used in IR spectrometer applications. Their temperature is constrained by their melting point and is typically regulated to 1050 °C.
Superior IR systems employ specialized ceramic sources operating between 1200 °C and 1400 °C, covering a broader spectral range. Their peak intensity can be up to twice as high; in spectroscopy, greater available light improves counting statistics and the signal-to-noise ratio.
The Need for More
The market demonstrates that fuel compositions are highly diverse. Though fuel technology continues to pursue greater combustion efficiency and higher energy output, environmental requirements call for low-emission fuels containing minimal quantities of environmentally problematic exhaust substances.
In addition, as global fuel prices rise, adulteration is no longer confined to particular regions and is becoming a worldwide issue. Modern fuel analyzers must accommodate this diversity through intelligent, flexible systems that automatically adjust to evolving fuel compositions.
Given the list of 7400 fuel additives, incorporating every substance into a model to represent every possible fuel type is clearly impractical.
Grabner Instruments’ new FTIR analyzer, the Grabner IRV Pro, is preconfigured to measure the most important fuel parameters and can learn to analyze additional substances from a simple concentration series.
It automatically establishes detection limits, while intelligent statistical algorithms help prevent calibration errors for new substances. This capability addresses the need to analyze more substances and also permits definition of typical contaminants.
Adding new and different substances carries the risk that their absorption may fall outside the conventional mid-IR range or be too strong or too weak. Extending the mid-IR range at both the low- and high-energy ends (especially adding the NIR range) allows the IRV Pro to overcome these limitations.
Substances with related molecular structures, such as alcohols, may also display overlapping bands that are difficult to separate when spectral resolution is insufficient. At the 4 cm-1 resolution typical of conventional FTIR instruments, methanol and ethanol bands are difficult to distinguish.
With resolution below 2 cm-1, the IRV Pro resolves this issue. Bands in a better-resolved spectrum are not only narrower but also taller, which helps lower detection limits for substances such as benzene.
Greater precision and stability of results are also needed. Although temperature does not strongly alter the IR spectra of solids or liquids, elevated temperatures increase molecular vibration and movement, broadening spectral bands. In other words, absorption bands widen as specimen temperature rises.
The IRV Pro features a unique double-cell spectrometer that maintains high precision regardless of the environmental conditions in which it operates. To prevent atmospheric CO2 or water vapor from interfering with measurements, the spectrometer is fully enclosed in a sealed, airtight housing filled with specially dried gas.

Figure 4. New IRV Pro from Grabner Instruments. Image Credit: Grabner Instruments
Conclusion
As more fuels must be analyzed within shorter timeframes and adulteration becomes an increasingly global concern, FTIR systems are receiving greater attention for general fuel screening outside conventional laboratories.
They are often the only screening instruments available at remote fuel terminals, small customs laboratories at borders, pipeline terminals, or mobile laboratories. As a result, there is growing demand to predict additional conventional fuel parameters. Beyond octane number and volatility characteristics, certain cold-flow properties, such as viscosity and flash point, may also be worth estimating.
A modern, highly flexible fuel analyzer such as the Grabner IRV Pro supports the development of new fuel models. As an additional feature, it incorporates a density meter. Density is an important fuel parameter that can be measured directly, and it also enables conversion of concentrations between mass percent and volume percent.
Because these instruments may be used in remote locations or mobile laboratories, greater physical durability is also required. The IRV Pro spectrometer is therefore housed in a hermetically sealed aluminum enclosure mounted on specialized suspensions within the instrument housing.
Durability requirements extend to the results as well. The stability of the measurement itself has been described, and prediction reliability is enhanced through use of the NIR region. This not only reduces noise but also allows property prediction calculations to benefit from the expanded spectral range.

This information has been sourced, reviewed, and adapted from materials provided by Grabner Instruments.