Why Materials With the Same MFR Can Behave Differently

The flow behavior of polymers used in the plastics industry is often measured via the melt flow rate (MFR). This metric is one of the simplest and most commonly used test methods for determining plastic flow behavior; however, it isn’t always good enough for determining the melt properties of a substance.

This article discusses a practical case comparing the analysis of two LDPE (low-density polyethylene) samples with the same MFR value, but exhibiting different behavior during processing.

Different suppliers produced both LDPE samples, though both samples had an MFR of 4.0 g per 10 minutes. This difference was significant, as one of the LDPEs caused substantial problems during production.

Materials and Methods

  • Polymer: LDPE

Test Arrangements

  • Torque-Rheometer System: HAAKE PolyLab OS Torque Rheometer (Thermo Scientific)
  • Torque-Rheometer drive unit: HAAKE RheoDrive 7 OS Drive Unit from (Thermo Scientific)
  • Analysis software: HAAKE PolySoft OS Capillary Software (Thermo Scientific)
  • Laboratory single screw extruder: Rheomex 19/25 OS
  • Extruder screw: L = 25 x D, compression ratio 2:1
  • Melt pump fitted with bypass valve
  • Melt-pressure sensors
  • Slit capillary die: 0.8 x 2.0 mm

Test Conditions

  • Liquid-cooled extruder feeding zone
  • Profile extruder temperature: 180 °C, 240 °C, and 280 °C
  • Melt pump temperature: 280 °C
  • Temperature at die: 280 °C
  • Extruder speed: 100 rpm
  • Melt pump speed: five to 60 rpm (programmed)

Measurement with melt-pump

Figure 1. Measurement with melt-pump. Image Credit: Thermo Fisher Scientific – Materials Characterization

Test Procedure

Polymer pellets were melted and homogenized in the extruder before being transferred into the melt pump via a bypass valve (Figure 1).

By adjusting the pump speed, a defined volume (Q) was driven through the slit-capillary die. This setup allows the apparent shear rate to be calculated directly from the capillary dimensions (W, H) and the measured volume flow (Q).

The pressure drop (p’) in the capillary is measured via pressure transducers, while the shear stress (τ) is calculated from the capillary geometry and the pressure drop.

It is then possible to calculate apparent viscosity (η) from the apparent shear rate and the shear stress.

Gradually increasing the melt pump’s speed increases shear rates. The viscosity function of the polymer melt is obtained by measuring the pressure drop at each speed step.

slit-capillary measurement with LOPE (Ml 4)

Figure 2. slit-capillary measurement with LOPE (Ml 4). Image Credit: Thermo Fisher Scientific – Materials Characterization

Results and Discussion

Figure 2 displays viscosity measurement results for the two LDPE samples in a single graph.

The samples show the same viscosity behavior at low shear rates, but shear rates varied considerably during processing. The graph shows distinct differences between the samples at higher shear rates.

Because the MFR method only measures one point in the viscosity behavior at low shear rates, this method alone could not show these differences.

Summary

As this test shows, the extruder capillary rheometer was able to differentiate the two samples. The MFR method failed to do so because it captures only a single viscosity point, offering insufficient insight at low shear rates.

HAAKE PolyLab OS Torque Rheometer

HAAKE PolyLab OS Torque Rheometer. Image Credit: Thermo Fisher Scientific – Materials Characterization

Acknowledgments

Produced from materials originally authored by Matthias Jährling from Thermo Scientific.

Image

This information has been sourced, reviewed and adapted from materials provided by Thermo Fisher Scientific – Materials Characterization.

For more information on this source, please visit Thermo Fisher Scientific – Materials Characterization.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Thermo Fisher Scientific – Materials Characterization. (2025, September 26). Why Materials With the Same MFR Can Behave Differently. AZoM. Retrieved on September 26, 2025 from https://www.azom.com/article.aspx?ArticleID=24692.

  • MLA

    Thermo Fisher Scientific – Materials Characterization. "Why Materials With the Same MFR Can Behave Differently". AZoM. 26 September 2025. <https://www.azom.com/article.aspx?ArticleID=24692>.

  • Chicago

    Thermo Fisher Scientific – Materials Characterization. "Why Materials With the Same MFR Can Behave Differently". AZoM. https://www.azom.com/article.aspx?ArticleID=24692. (accessed September 26, 2025).

  • Harvard

    Thermo Fisher Scientific – Materials Characterization. 2025. Why Materials With the Same MFR Can Behave Differently. AZoM, viewed 26 September 2025, https://www.azom.com/article.aspx?ArticleID=24692.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.