Water Determination in Vast Samples via KF Titration

Volumetric Karl Fischer (KF) titration is a fast, accurate technique for determining water content in a diverse range of samples. This determination is based on a chemical reaction involving the dissolution of water with a titrant containing iodine in an alcoholic solvent of specific composition.

The volumetric Karl Fischer method can be used with solid, liquid, and gaseous samples, provided water can be released from the sample into the Karl Fischer solvent.

The European Pharmacopoeia (Ph. Eur.) contains internationally recognized quality standards for analytical procedures that are used throughout the pharmaceutical industry. These standards considerably enhance the quality of pharmaceuticals produced, sold, and prescribed in Europe.

European Pharmacopoeia Chapter 2.5.12. (Water: Semi-micro determination) describes a suitability test for the volumetric Karl Fischer method for a defined combination of solvent, titrant, and sample.1,2,3

This suitability test is designed to verify KF titration systems in terms of linearity, recovery, and measurement range.

This article presents a suitability test for certified liquid water standards performed using a Mettler Toledo EVA V3 titrator.

Overview

The suitability test for the volumetric Karl Fischer technique contained in Chapter 2.5.12. of the European Pharmacopoeia was designed to verify the accuracy of Karl Fischer determinations for a known combination of titrant, solvent, and sample. This procedure is appropriate for corresponding sample sizes from 2.5 to five milligrams of water per injection.

Method A of Ph. Eur. 2.5.12. requires the water content of the sample under investigation to be determined in an initial sample injection. Five more sample injections are then performed.

Determined water content of each sample is then compared with the amount of water added, with the injection recovery of each also calculated.

A linear regression is done on the water content found in the cumulated water, which is determined by titration. This is compared with the cumulated added water contents following each sample.

The guidance provides a series of acceptance criteria for the determination to be found suitable:3

  • Between 97.5 % and 102.5 % mean recovery for found versus added water amounts
  • A slope between 0.975 and 1.025 on the linear regression line
  • Less than 2.5 % relative x- and y-axis intercept errors

In the example presented here, the EVA V3 KF Titrator was used to perform a suitability test in line with Ph. Eur. 2.5.12. This was done using a certified liquid water standard as the sample.

Procedures

  1. Begin method M903.
  2. Inject an initial sample as a blank to determine the water content. This should be done using the back-weighing technique to determine sample weight.
  3. Inject a further five samples with sample sizes of around 50 % to 100 % of the initial sample for the acceptance test. This should also be done using the back-weighing technique to determine the sample sizes.

Chemistry

ROH + SO2 + 3 RN + I2 + H2O (RNH)·SO4R + 2 (RNH)I

Solutions

  • Titrant: HYDRANAL Composite 5, c = 5 mg H2O/mL, one-component KF Titrant
  • Solvent: HYDRANAL Methanol dry
  • Standard: HYDRANAL Water standard 10.0, water content 10 mg/g = 1 %

Instruments and Accessories

  • Karl Fischer Titrator EVA V3 (30869281)
  • Analytical balance
  • 10 mL syringe and needle

Karl Fischer Titrator EVA V3 assembly

Figure 1. Karl Fischer Titrator EVA V3 assembly. Image Credit: Mettler-Toledo - Titration

Results

A liquid water standard with a certified water content of 10.03 mg/g was used to determine the titrant concentration before sample determination. The concentration was determined as 5.218 mg/mL (n=3).

Six injections of water standard 10.0 were recorded, with the first injection used to determine the water concentration.

The result was then treated as a blank value in later calculations of the amount of added water in relation to sample injections two to six. Table 1 summarizes these results.

Table 1. Found and added water contents for the KF suitability test according to Ph. Eur. 2.5.12. Source: Mettler-Toledo - Titration

Standard aliquot Sample size [g] Water found [mg] Content found [mg/g]
1 (Blank) 1.0374 10.34910 9.976
Standard aliquot Sample size [g] Water found [mg] Content found [mg/g]
2 0.6744 6.68240 6.72781
3 0.5276 5.20849 5.26334
4 0.6128 6.05345 6.11329
5 0.7264 7.21569 7.24657
6 0.7976 7.95276 7.95686

Linear regression and the defined acceptance parameters were automatically calculated using the instrument’s programmed method.

Table 2. Calculated test parameters for the KF suitability test according to Ph. Eur. 2.5.12. Source: Mettler-Toledo - Titration

Test parameter Found value
Mean percentage recovery 99.365 %
Slope of the linear regression line 0.994
y-Axis intercept percentage error -0.29 %
x-Axis intercept percentage error 0.28 %

All test parameters were found to be confidently inside the acceptance range defined by Ph. Eur. 2.5.12.

Remarks

This method complied with Ph. Eur. 2.5.12. Method A, with all parameters corresponding to Ph. Eur.’s acceptance criteria calculated by the instrument’s programmed titration method.

The first injection of the sample series was treated as a blank value to determine the sample’s water content. The determined water content was used to calculate the absolute amount of water added:

Water added = m • Content found.

Where m is the sample size measured in grams.

A new blank value is created in the setup to use the blank value in the presented method. The name of the blank value is “samplecontent” in this example, and its unit is mg.

It is important to note that it may be possible to omit the water content determination step when the water content is already known. If already known, users may skip the blank determination at the beginning of the task.

Waste Disposal and Safety Measures

Users should wear personal protection at all times while working in the laboratory, including lab coats, safety glasses, and gloves. Sample solutions should be disposed of as organic solvent waste after neutralization.

Measured Values

Titration curve obtained for the last sample

Figure 2. Titration curve obtained for the last sample. Image Credit: Mettler-Toledo - Titration

Table 3. Measured values for sample 5. Source: Mettler-Toledo - Titration

Time [s] Volume [mL] Measured value [μA] Water [mg] Drift [μg/min]
0 0 20.71952 0 90.3
1 0.000207 13.72729 0.00108 62.1
2 0.006095 7.236539 0.031802 281.8
3 0.029644 3.458559 0.154673 1548.1
4 0.075158 1.605391 0.392149 4533.6
5 0.139393 0.765484 0.727306 8871.1
6 0.213755 0.525218 1.115302 13523.7
7 0.298424 0.635714 1.557077 17714.5
8 0.377463 1.150055 1.969476 20771
9 0.457639 1.922473 2.387808 22534.6
10 0.533948 2.864141 2.785963 23398.6
53 1.527817 26.42984 7.97164 1944.1
54 1.529032 27.26243 7.977979 1752.6
55 1.529187 28.31202 7.978788 1238.7
56 1.529252 29.52222 7.979127 768.6
57 1.529471 30.55427 7.98027 484.2
58 1.529467 31.47755 7.980249 308.4
59 1.52947 32.18131 7.980265 187.3
60 1.530462 32.04423 7.985441 153.9
61 1.530468 32.13412 7.985472 155
62 1.530469 32.0181 7.985477 115.3
63 1.53047 32.02625 7.985482 69.9

Method

General Settings. Source: Mettler-Toledo - Titration

. .
Name M903
ID KFV Suitability Test
Compatibility Titration
Model compatibility V3
Method type KF Volumetric
SOP NO
Task comment NO

Configuration. Source: Mettler-Toledo - Titration

Analysis
Analyze more than one sample YES
Initial sequence NO
Final sequence YES
Determine blank YES
Open series NO
Analysis start Manual
Number of samples 5
Create statistics NO
Activate ‘KF conditioning’ after the method ends YES
KF cell KF cell 1
Category Volumetric (large)
Unit for drift μg/min
Work with solvent exchange YES
KF pump dPump KF 1
Start criteria Absolute drift values
Min. start drift 0.0 μg/min
Max. start drift 50.0 μg/min
Blank
Blank measurement Decide at task start
Open series NO
Number of blanks 1
Create statistics NO
Blank calculation (mean value) R1
Blank “B” samplecontent
Unit “B” mg
Result limits NO
Live View
Displayed results (Sample) 3
Field 1 R3: water found
Field 2 R4: water added
Field 3 R7: recovery
Displayed results (Blank) 1
Field 1 R1: water content
Displayed results (Final) 5
Field 1 R8: mean recovery
Field 2 R9: slope
Field 3 R12: R2
Field 4 R13: e1
Field 5 R14: e2
Analysis graph: Horizontal axis Time
Analysis graph: Vertical axis Measured value
Additional curve Drift
Conditioning graph: Horizontal axis Time
Conditioning graph: Vertical axis Drift

Blank Sequence. Source: Mettler-Toledo - Titration

1 Drift (Determination – online)
Drift determination YES
Determine online (during conditioning) YES
2 Blank (Addition)
Prompt for blank addition YES
Blank detection Automatic
Prompt for blank size YES
3 Titration (KF Vol)
Resources
Titrant Titrant Titrant 1
Nominal concentration 5 mg/mL
Sensor Sensor dSens M143
Category Polarized
Stirrer Stirrer Stirrer 1
Category Magnetic
Stir speed 35 %
Titration
Preparation Stir before titration 2 seconds
Control Control focus Accuracy
Indication Amperometric
Unit μA
Potential (Upol) 100 mV
Set current 30 μA
Cautious mode NO
Termination Type Drift stop relative
Drift relative 50.0 μg/min
Delay 0 seconds
Min. time 0 seconds
Max. time ∞ seconds
At Vmax 10 mL
4 Result R1: water content
Formula type Editable
Result name water content
Formula (VEQ*CONC-TIME*DRIFT/1000)/m
Unit mg/g
Decimal places 3
Result limits NO
5 Result R2: water found
Formula type Editable
Result name water found
Formula VEQ*CONC-TIME*DRIFT/1000
Unit mg
Decimal places 5
Result limits NO

Sample Sequence. Source: Mettler-Toledo - Titration

1 Drift (Determination – online)
Drift determination YES
Determine online (during conditioning) YES
2 Sample (Addition)
Prompt for blank addition YES
Blank detection Automatic
Prompt for blank size YES
3 Titration (KF Vol)
Resources
Titrant Titrant Titrant 1
Nominal concentration 5 mg/mL
Sensor Sensor dSens M143
Category Polarized
Stirrer Stirrer Stirrer 1
Category Magnetic
Stir speed 35 %
Titration
Preparation Stir before titration 2 seconds
Control Control focus Accuracy
Indication Amperometric
Unit μA
Potential (Upol) 100 mV
Set current 30 μA
Cautious mode NO
Termination Type Drift stop relative
Drift rel2ative 50.0 μg/min
Delay 0 seconds
Min. time 0 seconds
Max. time ∞ seconds
At Vmax 10 mL
4 Result R3: water found
Formula type Editable
Result name water found
Formula VEQ*CONC-TIME*DRIFT/1000
Unit mg
Decimal places 5
Result limits NO
5 Result R4: water added
Formula type Editable
Result name water added
Formula m*B
Unit mg
Decimal places 5
Result limits NO
6 Result R5: water found cumulated
Formula type Editable
Result name water found cumulated
Formula sum(R3.Result)+R2[1].Result
Unit mg
Decimal places 5
Result limits NO
7 Result R6: water added cumulated
Formula type Editable
Result name water added cumulated
Formula sum(R4.Result)
Unit mg
Decimal places 5
Result limits NO
8 Result R7: recovery
Formula type Editable
Result name recovery
Formula R3.Result/R4.Result*100
Unit %
Decimal places 3
Result limits NO

Final Sequence. Source: Mettler-Toledo - Titration

1 Result R8: mean recovery
Formula type Editable
Result name mean recovery
Formula mean(R7.Result)
Unit %
Decimal places 2
Result limits YES
Lower limit 97.5
Upper limit 102.5
Action if limits exceeded Continue task
2 Result R9: slope
Formula type Editable
Result name slope
Formula slope(R5.Result;R6.Result)
Unit
Decimal places 5
Result limits YES
Lower limit 0.975
Upper limit 1.025
Action if limits exceeded Continue task
3 Result R10: y-intercept
Formula type Editable
Result name y-intercept
Formula intercept(R5.Result;R6.Result)
Unit mg
Decimal places 5
Result limits NO
4 Result R11: x-intercept
Formula type Editable
Result name x-intercept
Formula -R10.Result/R9.Result
Unit mg
Decimal places 5
Result limits NO
5 Result R12: R2
Formula type Editable
Result name R2
Formula GoodnessOfFit(R5.Result;R6.Result)
Unit
Decimal places 5
Result limits NO
6 Result R13: e1
Formula type Editable
Result name e1
Formula 100*(R10.Result-R2.Result)/R2.Result
Unit %
Decimal places 2
Result limits YES
Lower limit -2.5
Upper limit 2.5
Action of limits exceeded Continue task
7 Result R14: e2
Formula type Editable
Result name e2
Formula 100*(abs(R11.Result)-R2.Result)/R2.Result
Unit %
Decimal places 2
Result limits YES
Lower limit -2.5
Upper limit 2.5
Action if limits exceeded Continue task
8 Report

References and Further Reading

  1. Angew. Chem., vol. 48, pp. 304 – 396, 1935
  2. Scholz, E. (1984). Karl Fischer TitrationChemical Laboratory Practice. Berlin, Heidelberg: Springer Berlin Heidelberg. https://doi.org/10.1007/978-3-642-69989-4.
  3. Linh. 2.5.12. Water semi-micro determination. (online) Available at: https://www.scribd.com/document/681905573/2-5-12-Water-semi-micro-determination.

Acknowledgments

Produced from materials originally authored by Mettler Toledo.

Image

This information has been sourced, reviewed, and adapted from materials provided by Mettler-Toledo - Titration.

For more information on this source, please visit Mettler-Toledo - Titration.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Mettler-Toledo - Titration. (2025, December 15). Water Determination in Vast Samples via KF Titration. AZoM. Retrieved on December 15, 2025 from https://www.azom.com/article.aspx?ArticleID=24754.

  • MLA

    Mettler-Toledo - Titration. "Water Determination in Vast Samples via KF Titration". AZoM. 15 December 2025. <https://www.azom.com/article.aspx?ArticleID=24754>.

  • Chicago

    Mettler-Toledo - Titration. "Water Determination in Vast Samples via KF Titration". AZoM. https://www.azom.com/article.aspx?ArticleID=24754. (accessed December 15, 2025).

  • Harvard

    Mettler-Toledo - Titration. 2025. Water Determination in Vast Samples via KF Titration. AZoM, viewed 15 December 2025, https://www.azom.com/article.aspx?ArticleID=24754.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.