Wastewater Metal Analysis Using ICP-OES Techniques

Assessing wastewater for metallic contamination is critical for safeguarding public health and the environment from the adverse effects of inadequately treated municipal wastewaters and specific industrial discharges.

 

Image Credit: Andromeda stock/Shutterstock.com

However, the regulatory frameworks for wastewater differ significantly between countries. In the United States, the Environmental Protection Agency (EPA), in conjunction with state agencies, negotiates discharge permits under the National Pollutant Discharge Elimination System (NPDES).

This process takes into account federal guidelines for industrial categories (40 CFR, Parts 405-471) and the ecological sensitivity of the receiving water bodies at each site.1

The treatment of wastewater within the EU is governed by the Urban Wastewater Treatment Directive (UWTD) and the Integrated Pollution and Control (IPC) Directive, both operating under the EU Water Framework Directive (WFD).

German wastewater regulations, known as the Abwasserverordnung (AbwV), integrate various EU directives pertaining to discharge limits for pollutants, aiming to reduce environmental pollution.

Unlike the wastewater discharge regulations enforced by the Clean Water Act (CWA) in the US, the German AbwV does not distinguish between direct and indirect discharges. Both categories must comply with the requirements set for individual industries or communal wastewater.

Consequently, wastewater requires the measurement of a diverse range of metals at varying concentrations, within different water matrices.

Several inorganic analytical methods are available for determining the elemental composition of wastewater. These include atomic absorption spectroscopy (AAS), inductively coupled plasma optical emission spectroscopy (ICP-OES), and ICP mass spectrometry (ICP-MS).

The most appropriate technique for each specific requirement can be chosen based on the number of elements and samples involved. While ICP-MS excels in highly precise trace detection, it is not well-suited for accurately and precisely determining mineral content at levels of 200 mg/L.

Standard ICP-OES instruments generally provide superior performance for assessing mineral and pollutant concentrations, but frequently lack the sensitivity needed to reliably detect toxic trace metals. The application of ICP-OES is described in various standard procedures, such as US-EPA Method 200.7 or ISO 11885.2-3

The US EPA developed Method 200.7 specifically for the determination of metals and trace elements in waters and waste materials using ICP-OES. This method is applicable to a variety of sample types, but it is extensively employed for wastewater applications.

The method incorporates a quality control program to ensure the proper operation of both the instrument and the analytical methodology during the analysis of wastewater samples.

In this study, the PlasmaQuant 9200 Elite ICP-OES - recognized for its high sensitivity, broad linear dynamic range, and exceptional spectral resolution - was evaluated for wastewater analysis.

The approach presented here details the analysis of pollutants and trace elements in wastewater using a standard ICP-OES configuration. The suitability of this methodology was validated through a quality control program aligned with Method 200.7.

This included: the analysis of several certified reference materials, participation in a national proficiency testing program for external quality assurance, assessment of spike recoveries in challenging matrices, and monitoring of the system's long-term stability.

Materials and Methods

Sample Preparation

All laboratory equipment was cleaned with deionized (DI) water from a PURELAB system (18.2 MΩ cm, ELGA LabWater, High Wycombe, England).

Single- and multi-element working standards were prepared by serial volume/volume dilution in polypropylene tubes from stock solutions (Merck, Sigma-Aldrich, CPAChem, Inorganic Ventures) using 1 % (v/v) sub-boiled distilled HNO3.

All samples (unless already stabilized) and blank solutions were acidified with HNO3 to yield a final acid concentration of 1 % (v/v). Yttrium (Y) was introduced online to all blanks, standards, and samples as an internal standard.

This study also incorporated various water types (surface and ground water) and wastewater samples, along with reference materials used for validating the developed method, as summarized in Table 1.

Three of the analyzed wastewater samples were part of a national round robin test (RRT) (“59th National Round Robin Test- Elements in Wastewater - 03/21” from the Staatliche sBetriebsgesellschaft für Umwelt und Landwirtschaft Sachsen(BfUL), Germany).

All samples were acidified with sub-boiled HNO3 to achieve a final acid concentration of 1 % (v/v), with the exception of the RRT samples.

Sample preparation for the RRT samples was conducted in accordance with EPA Method 3015A (SW-846) and DIN EN ISO 15587-2, which necessitate a microwave-assisted acid digestion step.

Consequently, a 25.0 (± 0.1) mL aliquot of the sample and 6.25 (± 0.10) mL of sub-boiled HNO3 were added to a digestion vessel (PM60). The mixture was gently swirled and allowed to stand for at least 15 minutes before the vessel was sealed.

Subsequent heating (20 minutes at 200 °C) was performed in a speedwave XPERT microwave digestion system. Afterward, the vessels were cooled to room temperature (RT) to prevent foaming and splashing. The resulting solutions were transferred to a graduated polypropylene tube and diluted to 50 mL with DI water.

In the case of Mercury (Hg), the sample preparation for the RRT samples was modified to fully comply with DIN EN ISO 12846: 2012-08 (E12). The samples arrived already stabilized (1 mL concentrated HCl per 100 mL) and were stored exclusively in glass containers throughout the entire procedure.

For preparation, 25 (± 0.1) mL of each sample were added to a 50 mL glass vial, combined with 0.5 mL of a 1:1 solution of potassium bromide (c (KBr) = 0.2 mol/L) and potassium bromate (c (KBrO3) = 0.033 mol/L), and incubated (24 hours at RT).

Following incubation, 10 µL of hydroxylammonium chloride (NH2OH x HCl, 120 g/L) and 0.5 mL of sub-boiled HCl were added to the reaction mixture, which was then analyzed immediately. Blank and standard solutions were prepared accordingly.

The determination of Hg is based on a cold vapor technique. The reducing agent comprised 3 g of sodium borohydride (NaBH4) and 1 g of sodium hydroxide (NaOH) dissolved in one liter of DI water, yielding concentrations of 0.3 % (m/v) and 0.1 % (m/v), respectively. Samples were diluted tenfold with 5 % (v/v) sub-boiled HCl and analyzed.

Table 1. List of samples and reference materials being analyzed. Source: Analytik Jena

Sample Supplier
Certified wastewater-Trace metals solution A (CWW-TM-A) High Purity Standards
Certified wastewater-Trace metals solution B (CWW-TM-B) High Purity Standards
Certified wastewater-Trace metals solution C (CWW-TM-C) High Purity Standards
Certified wastewater-Trace metals solution D(CWW-TM-D) High Purity Standards
ERM-CA713 Wastewater (trace elements) Sigma-Aldrich
RRT wastewater sample A BfUL
RRT wastewater sample B BfUL
RRT wastewater sample C BfUL
Industrial effluent – Inlet Automobile industry
Industrial effluent – Inlet Galvanic industry

 

Calibration

For each element, calibration levels were chosen based on expected concentration ranges. A minimum of three calibration standards were employed for each element, as seen in Table 2. Figure 1 depicts selected calibration curves.

Table 2. Concentration of calibration standards. Source: Analytik Jena

Standard Concentration [mg/L]
Ag, Al, As, B, Ba, Be,
Bi, Cd, Co, Cr, Cu, Fe,
Li, Mn, Ni, Pb, Se, Sr,
Te, Tl, V, Zn
Mg, P,
Si
Ca, K,
Na
Mo, Sb,
Sn, Ti
Hg
Calibration 0 0 0 0 0 0
Standard 1 0.01 0.01 - - -
Standard 2 0.1 0.1 0.1 - -
Standard 3 1.0 1.0 1.0 - -
Standard 4 - 10.0 10.0 - -
Standard 5 - - 100 - -
Standard 6 - - - 0.01 -
Standard 7 - - - 0.1 -
Standard 8 - - - 1.0 -
Standard 9 - - - - 0.005
Standard 10 - - - - 0.01
Standard 11 - - - - 0.025

 

Selected calibration functions

Selected calibration functions

Figure 1. Selected calibration functions. Image Credit: Analytik Jena 

Instrument Settings

The analysis was performed on a PlasmaQuant 9200 Elite ICP-OES. The sample introduction components and instrumental parameters were selected to achieve a high level of sensitivity for trace elements in matrices that might contain high salt concentrations.

A Teledyne CETAC ASX-560 Autosampler was used in conjunction with this instrument. An internal standard mixing kit was also employed to introduce a 10 mg/L yttrium internal standard solution online, resulting in an approximate final concentration of 1 mg/L. Table 3 provides a summary of individual settings and components.

Table 3. Instrument settings. Source: Analytik Jena

Parameter Specification
Plasma power 1250 W
Plasma gas flow 8.5 L/min
Auxiliary gas flow 0.4 L/min
Nebulizer gas flow 0.6 L/min
Nebulizer Concentric, SeaSpray, 2.0 mL/min, Borosilicate
Spray chamber Cyclonic spray Chamber, 50 mL, Borosilicate
Outer tube/Inner tube Quartz/Quartz
Injector Quartz, ID: 2 mm
Sample tubing PVC (red/red)
Internal standard tubing PVC (green/orange)
Pump rate 1.00 mL/min
Fast pump 4.0 mL/min
Measuring delay/Rinse time 45 s/15 s
Torch position 0 mm

 

Method and Evaluation Parameters

Table 4. Method parameters. Source: Analytik Jena

Element Line Plasma View Integration Read Time Evaluation
[nm]     [s] Pixel Baseline Fit Poly. Deg.
Y 371.030 axial/radial Peak 1/2 3 ABC1 auto
Ag 328.068 axial Peak 1 3 ABC auto
Al 394.401 axial Peak 3 3 ABC auto
As 188.979 axial Peak 10 3 ABC auto
B 249.773 axial Peak 3 3 ABC auto
Ba 455.403 radial Peak 1 3 ABC auto
Be 313.107 axial Peak 2 3 ABC auto
Bi 223.061 axial Peak 10 3 ABC auto
Ca 315.887 radial Peak 1 3 ABC auto
Cd 214.441 axial Peak 3 3 ABC auto
Co 228.615 axial Peak 3 3 ABC auto
Cr 267.716 axial Peak 1 3 ABC auto
Cu 327.396 axial Peak 3 3 ABC auto
Fe 259.940 axial Peak 1 3 ABC auto
Hg 184.886 axial Peak 10 3 ABC auto
K 766.491 radial Peak 1 3 ABC auto
Li 670.791 radial Peak 3 3 ABC auto
Mg 285.213 radial Peak 1 3 ABC auto
Mn 257.610 axial Peak 1 3 ABC auto
Mo 202.030 axial Peak 3 3 ABC auto
Na 589.592 radial Peak 1 3 ABC auto
Ni 231.604 axial Peak 3 3 ABC auto
P 177.436 axial Peak 3 3 ABC auto
Pb 220.353 axial Peak 10 3 ABC auto
Sb 217.581 axial Peak 10 3 ABC auto
Se 196.028 axial Peak 10 3 ABC auto
Si 251.611 radial Peak 1 3 ABC auto
Sn 189.927 axial Peak 3 3 ABC auto
Sr 407.771 radial Peak 1 3 ABC auto
Te 214.281 axial Peak 10 3 ABC auto
Ti 334.941 axial Peak 3 3 ABC auto
Tl 190.796 axial Peak 10 3 ABC auto
V 292.401 axial Peak 3 3 ABC auto
Zn 206.200 axial Peak 1 3 ABC auto

1 ... Automated Baseline Correction

Results and Discussion

EPA method 200.7 mandates a formal quality control (QC) program. This program involves, at a minimum, the initial demonstration of laboratory capability, and the ongoing analysis of blanks and other laboratory standards to ensure instrument performance.

The first demonstration of laboratory capability includes the determination of method detection limits (MDLs) and linear dynamic range (LDR), periodical checks of blank solutions, instrument performance checks (IPC), and spectral interference checks (SIC). Additionally, the accuracy and long-term stability of the method must be evaluated.

Linear Dynamic Range and Method Detection Limits

According to Method 200.7, the LDR is defined as the upper limit at which values are recovered within 10 % of the actual concentration when determined against the calibration curve used for analysis.

The MDLs are derived from seven measurements of a blank solution that has been spiked at a concentration two to three times the instrument detection limit. The standard deviation of these seven measurements is then multiplied by 3.14 (at a 99 % confidence level) to ascertain the MDL.

Table 5 presents the method-specific LDRs and MDLs.

Table 5. Method detection limits (MDL) and upper limit of the linear dynamic range (LDR) for the employed analytical lines. Source: Analytik Jena

Element Line
[nm]
Plasma
View
MDL
[μg/L]
LDR
[mg/L]
Ag 328.068 axial 0.14 20*
Al 394.401 axial 0.55 100*
As 188.979 axial 0.35 100*
B 249.773 axial 0.20 100*
Ba 455.403 radial 0.06 10
Be 313.107 axial 0.01 5
Bi 223.061 axial 0.35 100*
Ca 315.887 radial 1.38 250
Cd 214.441 axial 0.04 10
Co 228.615 axial 0.12 50
Cr 267.716 axial 0.10 50
Cu 327.396 axial 0.11 100*
Fe 259.940 axial 0.08 25
Hg 184.886 axial 0.14 10*
K 766.491 radial 22.0 200
Li 670.791 radial 0.23 25
Mg 285.213 radial 0.25 100*
Mn 257.610 axial 0.02 10
Mo 202.030 axial 0.13 100*
Na 589.592 radial 4.08 250
Ni 231.604 axial 0.12 20*
P 177.436 axial 1.25 100
Pb 220.353 axial 0.35 20*
Sb 217.581 axial 0.40 100*
Se 196.028 axial 1.25 100*
Si 251.611 radial 2.90 100
Sn 189.927 axial 0.25 100*
Sr 407.771 radial 0.04 5
Te 214.281 axial 0.88 100*
Ti 334.941 axial 0.02 20*
Tl 190.796 axial 0.45 100*
V 292.401 axial 0.04 50
Zn 206.200 axial 0.08 10

* Upper limit of test: even higher concentrations are possible by fulfilling the 90 % recovery criteria

Laboratory Reagent Blank (LRB) and Laboratory Fortified Blank (LFB)

Method 200.7 requires regular measurements of distinct blank solutions. One of these is the laboratory reagent blank (LRB), which is processed identically to samples, incorporating all reagents in the same volumes.

It should be analyzed with each batch of 20 or fewer samples (of the same matrix) to detect potential contamination from the laboratory environment and should not exceed 10 % of the determined analyte levels, or be smaller than 2.2 times the MDL.

The other blank is the laboratory fortified blank (LFB), which is prepared by spiking an aliquot of the LRB with a suitable analyte concentration. The LFB must also undergo the complete sample preparation process and should be analyzed with each batch of samples.

The accuracy, calculated as percent recovery, must fall within a ± 15 % control limit. All LRBs and LFBs were prepared and analyzed in compliance with Method 200.7 and met all stipulated criteria.

Quality Control Sample, Instrument Performance Check, and Stability

Immediately after calibration, two different QC samples must be analyzed to confirm the quality of calibration standards and instrument performance. These include the quality control sample (QCS) and the initial performance check (IPC) solution.

The IPC should originate from the same source, while the QCS must be prepared from a different stock solution. The recovery for both standards must be within ± 5 % of the specified value.

The specific standards were spiked at 0.01 mg/L (Hg), 0.5 mg/L (for most elements), 5 mg/L (for magnesium (Mg), phosphorous (P), and silicon (Si)), and 50 mg/L (for calcium (Ca), potassium (K), and sodium (Na)), respectively. Figure 2 indicates that all elements fulfilled the criterion.

Recoveries for the initial IPC (pink) and QCS (red).

Figure 2. Recoveries for the initial IPC (pink) and QCS (red). Image Credit: Analytik Jena

Furthermore, the quality program mandates the continuous measurement (after every 10 analyses and at the conclusion of the run) of the IPC throughout the entire analytical sequence.

The ongoing IPC results consistently remained within the acceptable range of 90–110 % of the known value, as shown in Figure 3. Figure 4 illustrates the behavior of the internal standard yttrium in both axial and radial observational views.

Relative standard deviations below 1.55 % signify highly stable instrumentation performance over the 14-hour measurement period.

Percentage recoveries in the IPC solutions throughout a 14-hour sequence. RSD values were below 1.65 % for all elements

Figure 3. Percentage recoveries in the IPC solutions throughout a 14-hour sequence. RSD values were below 1.65 % for all elements. Image Credit: Analytik Jena

Percentage recoveries of the internal standard (yttrium) throughout a 14-hour sequence. RSD values were below 1.55 % for both observational views.

Figure 4. Percentage recoveries of the internal standard (yttrium) throughout a 14-hour sequence. RSD values were below 1.55 % for both observational views. Image Credit: Analytik Jena

Spectral Interference Check

As a component of the QC program, a spectral interference check (SIC) solution must be tested periodically to verify the interelemental spectral correction routine.

For instruments that do not leverage interelement corrections (IEC), SIC solutions (containing similar concentrations of major sample components, e.g., ≥ 10 mg/L) can be used to confirm the absence of spectral interferences at the selected analytical wavelengths.

Given that the PlasmaQuant 9200 Elite uses a high-resolution optical system (2 pm @ 200 nm), access to the most common elemental lines is ensured, thereby rendering IEC tools obsolete. Nevertheless, an SIC solution containing 200 mg/L aluminum (Al) and 300 mg/L iron (Fe) was analyzed and seen to demonstrate no significant interferences at the chosen wavelengths.

Assessing Analyte Recovery

The chemical nature of the sample matrix can influence analyte recovery, meaning that Method 200.7 requires spiking experiments in at least 10 % of routine samples prior to sample preparation.

The recovery of the spiked analyte concentration must be within ±30 % of the concentration added to the sample. To address this, high-solid wastewater samples from the automobile and galvanic industries were spiked before sample preparation with 0.5 mg/L of the target elements.

Figure 5 presents the analyte recoveries, all of which fell within a ±20 % range, demonstrating the applicability of the analytical method.

Spike recoveries of 0.5 mg/L in wastewater samples originating from automobile (pink) and galvanic industry (red). Analytes were spiked before microwave digestion following EPA Method 3015A (SW-846) and DIN EN ISO 15587-2

Figure 5. Spike recoveries of 0.5 mg/L in wastewater samples originating from automobile (pink) and galvanic industry (red). Analytes were spiked before microwave digestion following EPA Method 3015A (SW-846) and DIN EN ISO 15587-2. Image Credit: Analytik Jena

Moreover, reference materials can be analyzed to show the validity of a method. As a result, four wastewater reference materials were analyzed. Table 6 lists the certified values and recovery rates for the tested reference materials.

All recoveries were within ±10 % of the certified value. Moreover, a wastewater reference material containing low concentrations - close to or under the legal limits of important drinking water regulations - of arsenic (As), mercury (Hg), and selenium (Se) was analyzed.

The recovery rates for these elements were within ±14 %, indicating the method's applicability for critical elements, such as As and Hg, even at lower concentration ranges that approach or fall below legal limits when using a standard sample introduction kit.

Table 6. Quantitative results for several certified reference materials (CRMs). Source: Analytik Jena

Element CWW-TM-A CWW-TM-B CWW-TM-C CWW-TM-D
Certified
[μg/L]
Recovery
[%]
Certified
[μg/L]
Recovery
[%]
Certified
[μg/L]
Recovery
[%]
Certified
[μg/L]
Recovery
[%]
Ag 10 97 50 96 150 99 250 98
Al 50 95 200 99 500 99 1000 99
As 10 95 50 97 150 97 250 96
B 50 100 200 101 500 102 1000 102
Baradial 50 99 200 98 500 99 1000 98
Be 10 95 50 97 150 98 250 98
Cd 10 98 50 97 150 97 250 98
Co 50 97 200 98 500 98 1000 98
Cr 50 98 200 100 500 100 1000 97
Cu 50 98 200 100 500 101 1000 101
Fe 50 98 200 100 500 99 1000 97
Mn 50 99 200 100 500 98 1000 98
Mo 50 97 200 98 500 98 1000 97
Ni 50 97 200 100 500 99 1000 98
Pb 50 99 200 100 500 97 1000 99
Sb 10 90 50 90 150 91 250 90
Se 10 94 50 100 150 100 250 101
Srradial 50 98 200 101 500 101 1000 99
Tl 10 94 50 100 150 102 250 99
V 50 98 200 101 500 100 1000 100
Zn 50 99 200 98 500 99 1000 99

 

Table 7. Quantitative results for certified reference material WW CA 713. Source: Analytik Jena

Element WW CA 713
Certified [μg/L] Recovery [%]
As 10.8 107
Cd 5.09 93
Cr 20.9 97
Cu 101 106
Fe 445 95
Hg 1.84 105
Mn 95 96
Ni 50.3 96
Pb 49.7 92
Se 4.9 86
Zn 78 100

 

External Quality Control: Round Robin Test

A round robin (or proficiency) test (RRT) is a method of external quality assurance for various measurement procedures.

Fundamentally, identical samples are analyzed using the same or different procedures in multiple laboratories. Comparing the results allows for statements to be made about general measurement accuracy or the measurement quality of the participating laboratory.

To demonstrate the performance of the PlasmaQuant 9200 and the applicability of the analytical method, the application laboratory at Analytik Jena's headquarters participated in a national round robin test.

This test involved the analysis of three wastewater samples. Sample preparation was conducted as described in the “Samples and Reagents” section. The organizer of the RRT specified the approved analytical methods and the corresponding regulations for sample preparation.

Table 8. Quantitative results for round robin test samples. Source: Analytik Jena

Element Wasterwater Sample A Wasterwater Sample B Wasterwater Sample C
Assigned
[μg/L]
Measured
[μg/L]
z-scorea Assigned
[μg/L]
Measured
[μg/L]
z-score Assigned
[μg/L]
Measured
[μg/L]
z-score
Al 1664.029 1640 -0.2 296.473 298 0.1 1256.799 1258 0.0
As 34.785 34.3 -0.1 86.134 81.4 -0.8 175.277 166.2 -0.9
Cd 2.741 2.69 -0.2 0.824 0.870 0.4 6.011 5.34 -1.4
Cr 385.734 415 1.6 96.992 92.3 -1.0 174.162 173 -0.1
Cu 258.785 257 -0.1 375.709 371 -0.2 86.059 84.8 -0.2
Fe 180.552 190 0.6 362.468 355 -0.7 850.259 841 -0.6
Hgb 0.371 0.299 -0.9 1.017 0.852 -0.7 1.522 1.28 -0.7
Ni 397.862 410 0.5 120.354 115 -0.8 199.558 188 -1.0
Pb 54.707 55.8 0.2 131.381 122 -0.7 76.425 71.6 -1.0
Zn 96.561 94.8 -0.2 167.401 156 -0.8 361.720 343 -0.8

a Calculation of z-score was performed in accordance to DIN 38402-45:2014-06
b Determined by using cold vapor ICP-OES

The concentration of Hg in the RRT samples was determined using cold vapor ICP-OES. To achieve this, the PlasmaQuant 9200 Elite was equipped with the hydride generation system HS PQ Pro (see Figure 6). With the addition of this system, the unique sensitivity of the HR-Array ICP-OES PlasmaQuant 9200 Elite was further improved, reaching detection limits below 10 ng/L for Hg.

Calibration curve for the analytical line of Hg at 184 nm using cold vapor ICP-OES (left). On the right, the HS PQ Pro is shown in the sample chamber of the PlasmaQuant 9200 Elite

Figure 6. Calibration curve for the analytical line of Hg at 184 nm using cold vapor ICP-OES (left). On the right, the HS PQ Pro is shown in the sample chamber of the PlasmaQuant 9200 Elite. Image Credit: Analytik Jena

Summary

The method presented here describes the use of a high-resolution ICP-OES in a standard configuration, fully adhering to the requirements of EPA Method 200.7 for a wide range of diverse wastewater samples.

It was demonstrated that the PlasmaQuant 9200 Elite successfully meets the stringent quality control requirements of the method by providing excellent sensitivity, accuracy, and long-term stability.

Furthermore, the instrument's performance and the applicability of the analytical method were successfully confirmed by passing a national round robin test.

PlasmaQuant 9200 Elite

Figure 7. PlasmaQuant 9200 Elite. Image Credit: Analytik Jena

The primary challenge in this application lies in analyzing elements across a broad concentration spectrum (from low µg/L to high mg/L) in a single run. Trace elements (e.g., As, Hg) need to be analyzed alongside highly concentrated minerals (e.g., Mg, Na) and pollutants (e.g., Al, Fe), a task successfully managed by the DualView Plus feature of the PlasmaQuant 9200 Elite.

Beyond the standard radial and axial plasma observation modes, it offers axial plus and radial plus views that attenuate the signal in the respective observation mode. The method described here leverages radial plasma observation for measuring high mineral levels, coupled with axial view for trace levels of toxic elements in a single measurement run. This eliminates the need for multiple dilutions to cover the entire concentration range.

The method validation encompassed accuracy determination on CRMs and several spiked wastewater samples, illustrating the suitability of the PlasmaQuant 9200 Elite ICP-OES system to meet wastewater and drinking water directives, such as the Safe Drinking Water Act, the European Drinking Water Directive, or the German drinking water ordinance.

A typical problem with ICP-OES instrumentation is insufficient sensitivity to comply with regulations for trace element detection. In this regard, the PlasmaQuant 9200 Elite offers high sensitivity, attributed to numerous technical features.

The high-efficiency generator produces a robust plasma, delivering strong and consistent signal intensity with minimal argon consumption, even for wastewater samples with high salt content.

Additionally, the spectral resolution of 2 pm @ 200 nm (Figure 8) makes the use of complex correction algorithms, such as inter-element correction, unnecessary. This enables reliable and interference-free quantification of trace elements by providing access to the most sensitive emission lines and high-definition peak shapes with enhanced sensitivity.

Comparison of PlasmaQuant 9200 (6 pm @ 200 nm) and PlasmaQuant 9200 Elite (2 pm @ 200 nm) illustrating the application advantage of high-resolution spectra (red: IPC, blue: blank, green: baseline correction (ABC)).

Figure 8. Comparison of PlasmaQuant 9200 (6 pm @ 200 nm) and PlasmaQuant 9200 Elite (2 pm @ 200 nm) illustrating the application advantage of high-resolution spectra (red: IPC, blue: blank, green: baseline correction (ABC)). Image Credit: Analytik Jena

The proposed setup fulfills all criteria for water and wastewater quality monitoring. The employment of the PlasmaQuant 9200 Elite allows water laboratories to conduct their entire elemental screening on a single instrument by using a standard sample introduction kit, leading to cost savings, reduced lab space requirements, and efficient use of time and labor.

Performance can be further augmented by incorporating a hydride system to lower the method detection limits for mercury and hydride-forming elements into the ng/L range.

Recommended Device Configuration

Table 9. Overview of devices, accessories, and consumables. Source: Analytik Jena

Article Article number Description
PlasmaQuant 9200 Elite 818-09201-2 High resolution ICP-OES
STANDARD KIT for PlasmaQuant 9x00 series 810-88006-0 Sample introduction kit for aqueous samples
Concentric nebulizer 2 mL/min
Borosilicate glass, for high salt content
418-13-410-609 Nebulizer for improved robustness and stability
Teledyne Cetac ASX-560 810-88015-0 Autosampler with integrated rinse function

 

References

  1. 40 CFR, Parts 405-471, U.S. Code of Federal Regulations
  2. EPA Method 200.7, “Determination of Metals and Trace Metals in Water and Wastes by Inductively Coupled Plasma Atomic Emission Spectrometry”, Rev. 4.4
  3. ISO. ISO 11885:2007. “Water quality - Determination of selected elements by inductively coupled plasma optical emission spectrometry (ICP-OES)”. Available at: https://www.iso.org/standard/36250.html.

Image

This information has been sourced, reviewed and adapted from materials provided by Analytik Jena.

For more information on this source, please visit Analytik Jena.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Analytik Jena. (2026, May 13). Wastewater Metal Analysis Using ICP-OES Techniques. AZoM. Retrieved on May 13, 2026 from https://www.azom.com/article.aspx?ArticleID=25271.

  • MLA

    Analytik Jena. "Wastewater Metal Analysis Using ICP-OES Techniques". AZoM. 13 May 2026. <https://www.azom.com/article.aspx?ArticleID=25271>.

  • Chicago

    Analytik Jena. "Wastewater Metal Analysis Using ICP-OES Techniques". AZoM. https://www.azom.com/article.aspx?ArticleID=25271. (accessed May 13, 2026).

  • Harvard

    Analytik Jena. 2026. Wastewater Metal Analysis Using ICP-OES Techniques. AZoM, viewed 13 May 2026, https://www.azom.com/article.aspx?ArticleID=25271.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.