Enhancing Microplastics Monitoring in Drinking Water

The measurement of microplastics (MPs) in drinking water is moving beyond its traditional role as a research application and becoming part of routine laboratory testing. As regulatory requirements develop, standardized monitoring programs are likely to increase the number of drinking-water samples that laboratories must process.

This creates a need for methods that combine dependable polymer identification with the speed and throughput required for large-scale analysis.

The LUMOS II ILIM, together with the ParticlePlus™ workflow, is designed to meet these requirements. Complete 25 mm filters can be analyzed with a total turnaround time of less than 30 minutes per sample. Automated measurement and AI-based polymer classification provide reproducible, operator-independent microplastic results suitable for routine laboratory use.

This article presents the method used to analyze a drinking-water sample collected in Hungary.

Experimental

ParticlePlus™ is a workflow developed for IR laser imaging (ILIM) microscopes. The study used the following configuration:

  • Sample details: A 20 L sample of raw drinking water was collected in Hungary and filtered through a 25 mm Whatman® aluminum oxide filter with a 0.2 μm pore size. The filter was supplied by Eurofins Analytical Services Hungary Kft.
  • Measurement setup: Automated whole-filter imaging was performed with the LUMOS II ILIM in transmission mode using the ParticlePlus™ software workflow. The laser wavenumber was stepped at intervals of 4 cm-1 across the MIR fingerprint range from 1800–1250 cm-1.
  • Data analysis: After acquisition, the hyperspectral dataset was automatically processed with Bruker’s Microplastic Identifier (MPID), without requiring operator input. The identification threshold was set at 750.

Results

The complete 25 mm filter required 17 minutes from the end of filtration to the final result. Full-filter acquisition of 22.5 million IR spectra took 12 minutes, while automated data processing required a further five minutes.

MPID carried out the microplastic analysis automatically using machine-learning (ML) models for unsupervised particle detection and classification. The software also determined particle-specific characteristics, including size, morphology, and spatial distribution. Further information about MPID is provided in Product Note M210.

In total, 71 particles were detected and identified. These included 59 polymer particles and 12 calcium carbonate (CaCO3) particles, demonstrating that the method can differentiate polymeric material from inorganic particles. Polyethylene (PE) was the most frequently detected polymer, with 21 particles, corresponding to 36% of all identified polymers.

The remaining polymer particles included polypropylene (PP, 11), polyethylene terephthalate (PET, 9), polyamide (PA, 7), and polymethyl methacrylate (PMMA, 1). Based on the volume of water analyzed, the measured concentration was 2.9 MPs per liter.

Detected particle sizes extended from 8 μm to 83 μm. The largest proportion, 43%, was within the 20–50 μm range. This result shows agreement with the particle-size range commonly reported for environmental microplastics in drinking water.

Comprehensive results summary for the drinking water sample

Figure 1. Comprehensive results summary for the drinking water sample. Image Credit: Bruker Optics

The performance of unsupervised polymer classification was demonstrated by comparing the mean spectra of the detected PE and PP particles, the two most abundant polymer groups, with their respective library reference spectra.

The strong agreement between the measured and reference spectra confirms the reliability of the AI-based identification and shows that accurate polymer classification can be achieved within a short analysis time.

Conclusion

  • A total analysis time of less than 30 minutes per 25 mm filter supports high-throughput routine testing.
  • MPID provides dependable particle detection and classification using unsupervised, AI-based analysis, eliminating the need for manual spectral review.
  • Polymer identity, particle size, morphology, and spatial distribution are determined simultaneously, providing detailed characterization of each particle.
  • The LUMOS II ILIM combined with ParticlePlu™ provides a suitable solution for large-scale drinking-water monitoring.

The Fastest IR Microscope Ever Built: LUMOS II ILIM

The Fastest IR Microscope Ever Built: LUMOS II ILIM

Image Credit: Bruker Optics

Acknowledgments

Produced using materials originally authored by Domenic Dreisbach (PhD), Product Manager, Bruker Optics GmbH & Co. KG.

Image

This information has been sourced, reviewed, and adapted from materials provided by Bruker Optics.

For more information on this source, please visit Bruker Optics.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Bruker Optics. (2026, November 03). Enhancing Microplastics Monitoring in Drinking Water. AZoM. Retrieved on November 03, 2026 from https://www.azom.com/article.aspx?ArticleID=25639.

  • MLA

    Bruker Optics. "Enhancing Microplastics Monitoring in Drinking Water". AZoM. 03 November 2026. <https://www.azom.com/article.aspx?ArticleID=25639>.

  • Chicago

    Bruker Optics. "Enhancing Microplastics Monitoring in Drinking Water". AZoM. https://www.azom.com/article.aspx?ArticleID=25639. (accessed November 03, 2026).

  • Harvard

    Bruker Optics. 2026. Enhancing Microplastics Monitoring in Drinking Water. AZoM, viewed 03 November 2026, https://www.azom.com/article.aspx?ArticleID=25639.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.