Why Use HR-ICP-OES for Trace Analysis of Lithium Carbonate?

This article is based on a poster originally authored by Sandro Spiller, Oliver Buettel, and Siqi Sun.

The potential of a new method for storing electrical energy, combined with increasing fossil fuel prices, offers electromobility the possibility of a long-term future in individual mobility. The future of electric drive systems could depend on batteries, considering the advantages of lithium-ion batteries (LIBs) like costs, weight, and energy density.

Requirements for LIBS are continually looking to increase operational lifetimes and improve capacity. The raw materials must be kept up by consistently providing higher quality.

Li2CO3, with a purity of 99.5%, is currently considered battery grade. The industry is searching for raw materials with purities equal to or greater than 99.9%. Demands for increased purity of the raw materials used in batteries mean that the analytical technique must guarantee sensitivity and matrix compatibility.

Download the Poster

Instrumentation

The inductively coupled plasma optical emission spectrometry (ICP OES) provides strong, rapid, multi-element analyses for various types of specimens. It is a favored instrument for finding trace impurities in lithium battery materials. For this examination, a high-resolution (HR) ICP-OES, the PlasmaQuant 9100 Elite, was utilized in trace elemental analyses in Li2CO3.

Extremely high resolution combined with exceptional plasma strength and increased instrument sensitivity provides the advantage of interference-free analyses of all the analytes examined in the low to sub-ppb range.

Excessive amounts of lithium ions in the plasma can create ionization interferences, resulting in false-positives in easily ionizable elements (EIE). This requires the calibration standards to be matrix-matched. To validate the analytical method’s suitability and assess instrument performance, analyte recovery, and extended stability tests were completed.

Instrument Settings

PlasmaQuant 9100 Elite ICP-OES completed the analyses. The instrument settings and the specimen introduction components were chosen to provide lower detection limits for trace elements in high salt-containing matrices.

The instrument had a glass concentric Seaspray nebulizer, a 2.0 mm inner diameter injector tube, and a cyclonic spray chamber with a dip tube installed. Increasing the space between the injector and plasma reduces the probability of salt deposits on the injector tip.

Sample Preparation

Working standards were formulated by serial volume/volume dilution in polypropylene of a multielement standard solution (Sigma Aldrich, 10 mg/L). All calibration standards and the blank solution (0.01 to 5.0 mg/L) were matrix-matched (5.0% (w/v) Li2CO3) and acidified (pH < 1) with concentrated (w ≥ 65%) nitric acid. A spiked (0.1 mg/L) blank solution was used as a specimen and for quality control.

The blank solution was prepared by dissolving 2.5 g of Li2CO3 in 10 mL DI water and adding 6 mL of concentrated nitric acid dropwise to prevent vigorous foaming. Lastly, the solution was filled to 50 mL with DI water and sonicated (5 min) to eliminate residual carbon dioxide.

Application advantage of high-resolution spectra (red: sample, black: spike, blue: blank, and baseline correction (green)). Interferences of the matrix components can be well separated from the analyte Ni (A and B) using HR Array optics (B). In contrast, employing instruments (A) with lower resolution results in spectral overlaps and broadening of signals.

Figure 1. Application advantage of high-resolution spectra (red: sample, black: spike, blue: blank, and baseline correction (green)). Interferences of the matrix components can be well separated from the analyte Ni (A and B) using HR Array optics (B). In contrast, employing instruments (A) with lower resolution results in spectral overlaps and broadening of signals. Image Credit: Analytik Jena US

Results and Discussion

The PlasmaQuant 9100 Elite provides exceptional sensitivity for many of the technical features. The four-wind coil combined with counter gas technology offers extended analytical zones in axial observation mode, resulting in as many signals as possible in ICP-OES.

The spectral resolution of 2 pm @ 200 nm (Figure 1) makes using difficult correction algorithms like the inter-element correction outdated.

This provided dependable quantification without the interference of trace elements by accessing the sensitive emission lines and highly visible peaks with exceptional sensitivity.

Table 1. Instrumental (IDL) and method detection limits (MLD) and spike recovery rates (0.1 mg/L) for certain elements in 5% (w/v) Li2CO3 solution. Source: Analytik Jena US

Element/Line IDL [μg/L] MDL [μg/kg] Spike Recovery
Al 396.152 0.9 0.018 99
Ca 315.887 1.1 0.022 97
Cu 327.396 0.5 0.010 100
Fe 259.940 0.3 0.006 99
K 766.481 0.7 0.014 99
Mg 280.271 0.1 0.002 100
Mn 259.372 0.1 0.002 99
Na 589.592 0.6 0.012 97
Ni 221.648 0.1 0.002 99
Pb 220.353 2.2 0.044 98
S 180.672 7.9 0.158 100
Si 251.611 1.8 0.036 99
Zn 206.200 0.2 0.004 99

 

Percentage recoveries of spiked analytes (0.1 mg/L) in a 5% (w/v) Li2CO3 solution throughout a 4.5-hour sequence. RSD values were below 2.2% for all elements.

Figure 2. Percentage recoveries of spiked analytes (0.1 mg/L) in a 5% (w/v) Li2CO3 solution throughout a 4.5-hour sequence. RSD values were below 2.2% for all elements. Image Credit: Analytik Jena US

Conclusions

The discussed technique meets the stability, sensitivity, and accuracy needed to complete the analyses of battery-grade Li2CO3 samples. The innovative torch design of PlasmaQuant 9100 Elite prevents salt deposits on vital glassware parts from forming and guarantees ease of torch handling throughout maintenance.

The PlasmaQuant 9100 Elite provides high sensitivity and the ability to dilute high salt-containing matrices, reducing downtime while meeting the examined product’s specific requirements.

Download the Poster

This information has been sourced, reviewed and adapted from materials provided by Analytik Jena US.

For more information on this source, please visit Analytik Jena US.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Analytik Jena US. (2024, April 16). Why Use HR-ICP-OES for Trace Analysis of Lithium Carbonate?. AZoM. Retrieved on April 30, 2024 from https://www.azom.com/article.aspx?ArticleID=23548.

  • MLA

    Analytik Jena US. "Why Use HR-ICP-OES for Trace Analysis of Lithium Carbonate?". AZoM. 30 April 2024. <https://www.azom.com/article.aspx?ArticleID=23548>.

  • Chicago

    Analytik Jena US. "Why Use HR-ICP-OES for Trace Analysis of Lithium Carbonate?". AZoM. https://www.azom.com/article.aspx?ArticleID=23548. (accessed April 30, 2024).

  • Harvard

    Analytik Jena US. 2024. Why Use HR-ICP-OES for Trace Analysis of Lithium Carbonate?. AZoM, viewed 30 April 2024, https://www.azom.com/article.aspx?ArticleID=23548.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.