Lab and Process Particle Analysis Systems

Dynamic Image Analysis

Dynamic image analysis involves the continuous streaming of particles through a measurement volume controlled by a camera. This free movement results in random orientations of the particles, allowing for accurate determination of their actual shape and size distribution from various perspectives.

Understanding the particle morphology provides valuable insights into the characteristic properties of the sample material, in addition to size.

Sympatec’s advanced evaluation software automatically generates all relevant size and shape characteristics following each measurement.

Source: Sympatec GmbH

Evaluation
Particle Size Equivalent sphere, equivalent perimeter, enclosing rectangle, Feret diameter, Chord length
Particle Shape Sphericity, aspect ratio, convexity, roundness
Fiber Characterization Fiber length, fiber diameter, volume-based fiber diameter, straightness, elongation
Volume models Sphere, ellipsoid, cylinder
Class limits Automatic generation depending on measuring range Intelligent, individual class limits generator ISO compliance by default

 

Dynamic Image Analysis in Lab and Process

The modular image analysis sensor QICPIC integrates size and shape analysis of disperse particle systems within a size range of less than 1 µm to 34,000 µm. Key features include speed, reproducibility, and comparability of measurement results with optimized sample quantities.

Integrated image analysis sensors PICTOS, PICTIS, and PICCELL facilitate process-related particle size and shape characterization, covering a size range from 1 µm to 10,000 µm.

Integration into control systems or automated laboratories is achieved using standard communication protocols such as Modbus®, Profibus®, and OPC (Open Platform Communications) with TCP/IP, RS232, or RS485.

Dispersion and Dosing

As each particle's image is evaluated separately, the quality of particle singulation is of utmost importance. The QICPIC's modular design allows for the application of dosing and dispersing units suitable for wet and dry products, providing flexibility to accommodate powders, granules, fibers, suspensions, and emulsions. This ensures product-specific characterization that aligns with the true nature of the application.

Source: Sympatec GmbH

Specifications | Quality of measurement results
Sensor QICPIC/L02 QICPIC/L06
0.55 - 34,000 μm
Frame rate | fps 225 @ 1.5 MP 500 @ 1.5 MP
Data rate 5 GBit/s 10 Gbit/s
plus PICTOS, PICTIS & PICCELL for process
Principle Pulsed LED
Range Optical modules M3 (10:1) up to M9 (1:3)
Results Accuracy σ < 1 % mean SD
Repeatability σ < 0.2 % narrow distribution
  σ < 1 % wide distribution
Comparability σ < 2 % mean relative SD
  |Δx50| < 5 % max. relative deviation
Software PAQXOS Control and evaluation software
Quality
assurance
Certification Standardized test procedure with reference materials
Validation According to FDA regulations
Conformity ISO 13322-2 | FDA 21 CFR 11

 

Image

This information has been sourced, reviewed and adapted from materials provided by Sympatec GmbH.

For more information on this source, please visit Sympatec GmbH.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Sympatec GmbH. (2025, October 15). Lab and Process Particle Analysis Systems. AZoM. Retrieved on October 15, 2025 from https://www.azom.com/article.aspx?ArticleID=24595.

  • MLA

    Sympatec GmbH. "Lab and Process Particle Analysis Systems". AZoM. 15 October 2025. <https://www.azom.com/article.aspx?ArticleID=24595>.

  • Chicago

    Sympatec GmbH. "Lab and Process Particle Analysis Systems". AZoM. https://www.azom.com/article.aspx?ArticleID=24595. (accessed October 15, 2025).

  • Harvard

    Sympatec GmbH. 2025. Lab and Process Particle Analysis Systems. AZoM, viewed 15 October 2025, https://www.azom.com/article.aspx?ArticleID=24595.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.