Advancing EDS Analysis with Next-Generation Detector Technology

insights from industryDr. Sebastian Schmidt & Dr. Purvesh SoniBruker Electron Microscope Analyzers

In this interview, AZoMaterials speaks with Dr. Sebastian Schmidt, Product Manager EDS at Bruker Electron Microscope Analyzers, and Dr. Purvesh Soni, Senior Application Scientist, about advances in EDS analysis and how segmented detector technology can deliver faster measurements, improved spectral resolution, and new opportunities for analyzing increasingly complex materials.

Can you please introduce yourself and your role at Bruker?

Sebastian: I am Sebastian Schmidt, a physicist who has been with Bruker since 2022. I currently work as a Product Manager for EDS on SEM where we have developed two new advanced EDS detectors that are changing what is possible in EDS. These are the ultra-fast and high-resolution XFlash® 7200 FIRE and the XFlash® FlatQUAD 2L.

Purvesh: My name is Dr. Purvesh Soni. I am a Senior Applications Scientist for EDS at Bruker in Berlin, where I work across a range of EDS applications. Recently my focus has been on developing new EDS applications enabled by the leap forward that this exciting new multi-segment detector technology brings. 

What challenges in modern EDS analysis led Bruker to develop its latest detector technology?

Sebastian: People using EDS on SEM are studying smaller and smaller structures while, at the same time, working with increasingly complex materials. EDS is also becoming more interesting for industrial applications, particularly for quality assurance.

One of the big goals is therefore to make measurements faster. In industrial applications, but also in scientific laboratories, you want to get to your results more quickly, save time, and increase productivity.

Traditionally, one way to improve measurement speed has been to increase detector area. However, conventional EDS detectors suffer from two fundamental physical limitations when you do that. First, when you increase the detector area, fewer X-rays can actually be processed in a given time. Second, system noise increases with detector volume, which has a negative effect on the spectral resolution of the EDS spectrometer.

With the XFlash® 7200 FIRE, we achieve a very large detector area of 200 mm2 by combining four smaller individual segments. Each segment has dedicated pulse-processing electronics, allowing us to acquire and process X-rays in parallel. We already use a similar concept with the XFlash® FlatQUAD 2L, but this is the first inclined detector featuring this technology.

As materials and devices become more complex and features get smaller, how are the requirements for EDS analysis changing?

Sebastian: Semiconductor samples are a good example. You need a high-resolution image or map of the elemental composition, but at the same time some elements are difficult to distinguish within an EDS spectrum. That means you also require very high spectral resolution.

If you simply increase detector area using a conventional design, you normally sacrifice spectral resolution. With four individual segments, we can achieve a larger detector area while maintaining excellent spectral resolution, yielding very good peak separation for almost all elements.

Measurement speed is also extremely important for spatial resolution. The longer you measure, the more sample drift you can experience, which can blur the image. Short measurement times are therefore key to obtaining high spatial resolution.

High-resolution SEM-EDS mapping reveals sub-5 nm chemical features within a FinFET fin structure, including an apparent W-penetration into the adjacent Ti region. Strongly overlapping Si-K, W-M and Hf-M X-ray lines are deconvoluted automatically, providing insight into layer conformity, interface integrity and local material penetration or intermixing relevant to semiconductor process development. Image Credit: Bruker Nano

What makes the XFlash® 7200 FIRE different from conventional EDS detector designs?

Sebastian: What really makes this detector special is its four individual-segment design rather than a single large SDD detector. This gives us a higher measurement speed because more X-ray signals can be detected and processed at a given time. At the same time, we achieve excellent spectral resolution.

It is this combination of measurement speed and spectral performance that makes this detector stand out compared with conventional designs.

For somebody using the microscope, this translates into higher productivity, particularly in quality assurance, but also in a conventional scientific laboratory. You get your results more quickly and can obtain more accurate results because you acquire more X-rays and therefore improve your counting statistics.

Why is this technology particularly useful for semiconductor EDS analysis?

Sebastian: We believe it is ideal for semiconductor analysis because these samples typically require very high spatial and spectral resolutions.

You want to detect as many of the X-rays emitted from the sample as possible to make an accurate analysis of the spatial elemental composition. In semiconductors, you also have the issue that many X-ray peaks overlap. With this detector, it is possible to distinguish the relevant peaks encountered in semiconductor science and industry.

It is also windowless, which is particularly important for samples containing light elements or when you need to work in the low-energy region using a low microscope acceleration voltage. Without a detector window, more of those low-energy X-rays can be detected.

There are also applications where you previously had to compromise between speed, sensitivity, and data quality. When measuring the small dimensions in semiconductors, a short measurement time is key in order to avoid sample drift affecting the spatial resolution. Also beam-sensitive and biological samples, for example, need to be measured in a very short time because prolonged exposure can damage the sample. You therefore need to collect as many emitted X-rays as possible within that limited measurement period.

The detector provides a very high solid angle, which, combined with four-channel parallel signal processing, makes it well-suited to applications where measurement time must be kept short.

SEM-EDS elemental map of a ground polymetallic ore acquired with the XFlash® FlatQUAD 2L detector (left), alongside the simultaneously acquired BSE image (right). Compositional EDS mapping with BSE imaging enables rapid correlation of elemental distributions with Z-contrast and particle morphology, supporting efficient identification and characterization of complex multiphase materials. The highly detailed elemental map was acquired in just 5 minutes at 15 kV, with an output count rate of 746,400 cps. Image Credit: Bruker Nano

How does the XFlash® FlatQUAD approach complement the XFlash® 7200 FIRE for challenging EDS analysis?

Sebastian: The XFlash® FlatQUAD 2L has a very different geometrical arrangement, compared with conventional detectors positioned at the side of the sample. The XFlash® FlatQUAD itself is positioned between the sample and the electron microscope pole piece.

There is an aperture in the XFlash® FlatQUAD 2L that allows the electron beam to reach the sample, while X-rays emitted by the sample are detected by an annular array of SDD detectors. It is also segmented into four sections, with each segment equipped with its own signal processor.

Because the detector is very close to the sample, you can achieve extremely high measurement speeds.

The geometry is also ideal for highly topographical samples because you have a very high take-off angle for the emitted X-rays. This minimizes shadowing. You can look into deep troughs and obtain elemental information from areas where conventional detector geometries may not be able to detect the emitted X-rays.

How can combining EDS analysis with backscattered electron imaging improve analytical workflows?

Sebastian: In some industrial applications, users first employ the backscattered electron (BSE) signal to identify particles or other features of interest. BSE imaging is fast, so once those features have been identified, you can then use EDS to determine their elemental composition.

This makes the process very efficient; you can identify the features of interest quickly and then use the XFlash® FlatQUAD 2L to determine their composition.

Purvesh: With BSE detection enabled alongside the XFlash® FlatQUAD 2L, you can also see material contrast very easily, which is useful when materials of different densities are present, as it helps you navigate rapidly to the desired region of interest.

Another benefit concerns charging samples. Charging effects can be more pronounced when using an SE detector, whereas with BSE imaging those effects are much lower.

SEM-EDS elemental map of a dental biofilm growing on human dentin, acquired with a low beam current and the XFlash® FlatQUAD 2L detector at 5 kV for 10 minutes with an output count rate of 198,500 cps.  This 3D structure, with dentin’s deep tubules and an intricate biofilm network, is highly challenging for EDS analysis. The XFlash® FlatQUAD 2L’s unique annular design minimizes shadowing effects to deliver a chemically resolved view of this complex biological system, rather than just the surface image that a conventional detector would provide. Image Credit: Bruker Nano

Which applications stand to benefit most from these advances in EDS analysis?

Sebastian: Semiconductors are a very good application for the XFlash® FlatQUAD 2L because of its extremely high measurement speed. Biological samples and highly topographical materials can also benefit significantly. There is also potential for studying biological cells and possibly investigating chemistry inside cells during different stages of growth.

Purvesh: We have worked with life science, semiconductor, and battery samples that require very low acceleration voltages. With a regular inclined EDS detector, count rates under these conditions can become extremely low. With the XFlash® FlatQUAD 2L, you can collect much higher count rates even at low acceleration voltages and with only a small probe current. That really changes the way you can perform EDS analysis.

What do these detector advances ultimately mean for scientists and engineers using an SEM day to day?

Sebastian: The main value is productivity. Scientists and engineers can make better use of their microscope time, perform more accurate analyses, and potentially achieve higher spatial resolution when the system is set up correctly.

Perhaps more importantly, these technologies may provide access to applications that were not accessible with conventional detectors. That is something we are still exploring.

Purvesh: With the very high collection efficiency of segmented detector technology, we can perform EDS analysis at extremely low acceleration voltages, for example around 500 V or 1 kV. In the past, at voltages such as 1 or 2 kV, users might primarily have performed imaging because EDS measurements were so difficult. These new detector technologies are opening new horizons for analytical work at those very low voltages.

Ultra-low-kV SEM-EDS elemental map of S. turris, a microscopic diatom, acquired using the XFlash® 7200 FIRE at 1.7 kV with an output count rate of 19,200 cps. At 1.7 kV, the Si Kα line is not excited, enabling chemical mapping of the nanostructured silica frustule using the ultra-low-energy Si L emission at 91 eV. The superior count statistics and sensitivity of the XFlash® 7200 FIRE to ultra-low-energy X-rays enable high-spatial-resolution elemental mapping under low-kV conditions well suited to beam-sensitive biological samples. Image Credit: Bruker Nano

Where do you see EDS detector technology developing next?

Purvesh: Bruker is known for innovation in this area, and I think the four-segment design represents a significant step in detector technology.

Previously, we had single silicon drift detector chips. Now, with four segmented chips, we can achieve very fast X-ray collection and processing, very low dead times, and at the same time very good spectral quality.

Sebastian: Looking forward, segmented EDS detector technology gives us the opportunity to optimize the collection solid angle even further and tailor measurement speed to particular applications, SEMs, or microscope configurations. There is a high degree of freedom in how this technology can be utilized, and that creates opportunities to extend EDS analysis into areas where analytical measurements have previously been difficult.

About the Speakers

Dr. Sebastian Schmidt is Product Manager for EDS on scanning electron microscopes & Software at Bruker Nano in Berlin. He received his PhD from the Helmholtz-Zentrum Berlin für Materialien und Energie and the Technical University Berlin for the investigation of grain boundaries in thin film semiconductors using TEM. Dr. Schmidt has contributed to the development, introduction, and communication of new products including the XFlash® FlatQUAD 2L and XFlash® 7200 FIRE. Bruker describes the latter as a next-generation inclined EDS detector designed for ultra-fast, high-resolution elemental analysis on SEM.

Dr.-Ing. Purvesh Soni is a Senior Applications Scientist at Bruker Nano in Berlin, specializing in SEM-EDS and advanced X-ray microanalysis. He earned his PhD from RWTH Aachen University and Max Planck Institute for Sustainable Materials in Düsseldorf, formerly MPIE, with a focus on electron microscopy and nanoscale characterization of semiconductor materials. At Bruker, he specializes in advanced EDS detector technologies, including the XFlash® 7200 FIRE and XFlash® FlatQUAD 2L detectors for advanced applications in semiconductor analysis, battery research, and life sciences. He focuses on ultra-low-kV microanalysis and the translation of new detector technologies into practical materials-characterization solutions.

Image

This information has been sourced, reviewed, and adapted from materials provided by Bruker Nano GmbH – Electron Microscope Analyzers.

For more information on this source, please visit Bruker Nano GmbH – Electron Microscope Analyzers.

Disclaimer: The views expressed here are those of the interviewee and do not necessarily represent the views of AZoM.com Limited (T/A) AZoNetwork, the owner and operator of this website. This disclaimer forms part of the Terms and Conditions of use of this website.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Bruker Nano GmbH – Electron Microscope Analyzers. (2026, November 26). Advancing EDS Analysis with Next-Generation Detector Technology. AZoM. Retrieved on November 26, 2026 from https://www.azom.com/article.aspx?ArticleID=25534.

  • MLA

    Bruker Nano GmbH – Electron Microscope Analyzers. "Advancing EDS Analysis with Next-Generation Detector Technology". AZoM. 26 November 2026. <https://www.azom.com/article.aspx?ArticleID=25534>.

  • Chicago

    Bruker Nano GmbH – Electron Microscope Analyzers. "Advancing EDS Analysis with Next-Generation Detector Technology". AZoM. https://www.azom.com/article.aspx?ArticleID=25534. (accessed November 26, 2026).

  • Harvard

    Bruker Nano GmbH – Electron Microscope Analyzers. 2026. Advancing EDS Analysis with Next-Generation Detector Technology. AZoM, viewed 26 November 2026, https://www.azom.com/article.aspx?ArticleID=25534.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.