How to Perform Rapid Water Analysis to Determine Trace Elements

In order to devise an approach to identify around 30 different metals and trace elements in waters, wastewaters and solid wastes by ICP-OES, the U.S. EPA produced Method 200.7.1

Established in 1994, Method 200.7 has been routinely performed in environmental labs across the United States.

With the continual expansion of industry, pollution and wastes are persistently produced in larger quantities, leading to an increasing number of samples, which necessitate analysis per Method 200.7.

In an attempt to meet the needs of increasing demand, PerkinElmer’s Avio® 560 Max entirely simultaneousICP-OES includes an integrated High Throughput System (HTS) for high-speed sample-to-sample times.

The HTS is a valve-and-loop system that significantly reduces the time needed for the sample to reach the nebulizer and washout time post-analysis by utilizing a vacuum to both quickly fill and wash the sample loop.

This work outlines the analysis of wastewaters following Method 200.7 utilizing the Avio 560 Max ICP-OES and advancing on previous work.2

Experimental

Samples and Sample Preparation

Development and evaluation of the methodology were achieved with four separate wastewater certified reference materials:

  • Wastewaters C, D and H (High Purity Standards™, Charleston, South Carolina, USA)
  • WasteWatR™, Trace Metals (ERA, Golden, Colorado, USA)

These standards were provided as concentrates and diluted in accordance with the certificate instructions. Samples were prepared in line with Method 200.7: 50 mL were acidified with 2 mL of 1:1 HNO3 and 1 mL of 1:1 HCl, then stationed in a hot block and heated to a temperature of ≈ 85°C.

The solutions were taken out with ≈ 20 mL final volume for cooling cool and using deionized water diluted to 50 mL for analysis.

All measurements were taken in opposition to external calibration curves, where the concentrations in Table 1 in 2% HNO3 (v/v) set the standards for preparation.

Table 1. Calibration Standards. Source: PerkinElmer, Inc.

Element Standard 1
(mg/L)
Standard 2
(mg/L)
Ag, Al, As, B, Ba, Be, Cd, Co, Cr, Cu, Fe, Li, Mn, Mo, Ni, P, Pb, Sb, Se, Si, Sn, Sr, Ti, Tl, V, Zn 0.5 1
Na, Mg, K, Ca 10.5 21

 

After preparation in 2% HNO3(v/v), the internal standard (yttrium, Y) was added to the sample stream via a port in the HTS valve. Interelement corrections (IECs) were administered to all measurements.

As detailed in Method 200.7, the IEC solutions were run at the appropriate concentrations, except where they were beyond the linear range. In these instances, the concentrations of the IEC solutions were modified to come within the linear range.

Instrumental Conditions

All analyses were conducted on an Avio 560 Max ICP optical emission spectrometer (OES), utilizing the High Throughput System (HTS)3 for the introduction of samples via an S23 Autosampler.

Parameters and conditions of the instrument are displayed in Table 2, along with the analytical wavelengths, analytes and plasma view modes in Table 3.

Table 2. Avio 560 Max ICP-OES with HTS Instrumental Parameters. Source: PerkinElmer, Inc.

Paramenter/Component Value/Description
Sample Uptake Tubing Black/Black (0.76 mm id) PVC
Internal Standard Tubing Green/Orange (0.38 mm id), PVC
Drain Tubing Gray/Gray (1.30 mm id), Santoprene
Nebulizer MEINHARD® K1
Spray Chamber Baffled Glass Cyclonic
Carrier 2% HNO3 (v/v)
Carrier Flow Rate 0.8 mL/min
Sample Loop Volume 1 mL
Injector 2.0 mm id Alumina
Nebulizer Gas Flow 0.70 L/min
Auxiliary Gas Flow 0.2 L/min
Plasma Gas Flow 8 L/min
Torch Depth -3
Integration Auto
Read Time Range 0.5-5 sec
Loop Fill Time 4 sec
Loop Rinse Time 3 sec
Replicates 2

 

Table 3. Analytes, Wavelengths, and Plasma View Mode. Source: PerkinElmer, Inc.

Element Wavelength (nm) Plasma View Mode
Ag 328.068 Axial
Al 308.215 Radial
As 188.979 Axial
B 249.677 Axial
Ba 493.408 Radial
Be 313.107 Radial
Ca 315.887 Radial
Cd 214.440 Axial
Ce 413.764 Axial
Co 228.616 Axial
Cr 267.716 Axial
Cu 324.752 Axial
Fe 238.204 Radial
K 766.490 Radial
Li 670.784 Radial
Mg 285.213 Radial
Mn 257.610 Axial
Mo 203.845 Axial
Na 589.592 Radial
Ni 231.604 Axial
P 178.221 Axial
Pb 220.353 Axial
Sb 206.836 Axial
Se 196.026 Axial
Si 251.611 Radial
Sn 189.927 Axial
Sr 421.552 Radial
Ti 334.940 Axial
Tl 190.801 Axial
V 292.402 Axial
Zn 206.200 Axial
Y (Internal Standard) 371.029 Axial and Radial

 

Conventional sample introduction components and conditions were utilized, including a complete argon flow of 9 L/min. While Method 200.7 asserts that four replicates are necessary, a typical implementation is the use of two or three replicates to improve sample throughput.

To ensure this is more relevant and works for commercial labs, all measurements were taken using two replicates. All analyses were conducted using auto-integration with a read-time range of 0.5-5 seconds, which yielded rapid analysis for high-level analytes while also permitting accurate measurements of low-level analytes.

To streamline analysis, auto background correction was implemented. Under these conditions, sample-to-sample time was in the region 60 seconds.

For optimum performance, auto integration with variable read times was used, which resulted in the sample-to-sample time varying depending on the analyte concentrations in the samples: for higher concentrations, sample-to-sample time will be under 60 seconds; for samples with lower concentrations, the sample-to-sample time can be just over 60 seconds.

To meet conventional operating procedures, three or four replicates can also be applied with minimal sample-to-sample time gain.

Results and Discussion

To ensure compliance with Method 200.7, quality control (QC), the appropriate criteria for sample handling and preparation as well as the instrumental analysis must be satisfied.

The following criteria specifically relates to instrumental analysis and will be evaluated to establish the validity of the Avio 560 Max system: instrument performance checks (IPCs), linear dynamic range, spectral interference checks (SICs), method detection limits (MDLs), stability and accuracy.

Using Smart Software to Facilitate Data Analysis

Syngistix for ICP software (version 5.1 or higher) incorporates many smart features to enable data analysis and interpretation, including smart methods, smart workflow, smart data and smart monitoring.

While all of these were utilized when receiving data for Method 200.7, only a small number of the features will be outlined here.

One of the most crucial aspects of obtaining solid data is reviewing and satisfying good linearity for calibration curves. This can be completed with ease using Data Viewer, as shown in Figure 1.

Calibration information displayed in Data Viewer.

Figure 1. Calibration information displayed in Data Viewer. Image Credit: PerkinElmer, Inc.

When data is obtained using Data Viewer, thumbnails of the calibration curves emerge, which give the user a brief overview. For more in-depth analyses, details of the calibration curve materialize when the thumbnails are clicked.

This shows the correlation equation, coefficients, entered concentrations, corrected intensities and calculated concentrations for each standard, as well as the residual error for each standard.

If modification of the calibration curve is required (for example, if a standard was out of range), the Edit Calibration window displays the same information (Figure 2), which enables editing of the calibration information.

Detailed calibration information displayed in the Edit Calibration window.

Figure 2. Detailed calibration information displayed in the Edit Calibration window. Image Credit: PerkinElmer, Inc.

When running long-term stability, it is also vital to monitor the stability of the internal standard.

The user is able to monitor the real-time performance of an active run due to the normalization of internal standard response (normalized to the calibration blank) displayed automatically in Data Viewer in real-time (Figure 3).

Internal standard plot in Data Viewer.

Figure 3. Internal standard plot in Data Viewer. Image Credit: PerkinElmer, Inc.

Initial QC: Initial Performance Check and Quality Control Sample

To substantiate the calibration curve’s quality, an initial performance check (IPC) must be analyzed directly after the calibration. The IPC must be a separate standard made equivalent to the stock standard as the calibration standards.

Additionally, it is necessary to make a quality control sample (QCS) to the same concentration as the IPC, but from a second-source standard to verify the concentration of the stock solutions utilized for the calibration standards.

Recover within 5% of their true values for both the IPC and QCS are required for the analysis to proceed.

For this assessment, all elements were present at 0.75 ppm, with the exception of the minerals (Ca, K, Mg, Na), which were added at 21 ppm. Figure 4 demonstrates that both the QCS and IPC recovered within the 5% needed for all analytes.

Recoveries for initial IPC and QCS.

Figure 4. Recoveries for initial IPC and QCS. Image Credit: PerkinElmer, Inc.

Method Detection Limits

Method 200.7 details that method detection limits (MDLs) are established by measuring a standard seven times which has been contaminated with analytes at 2-3 times the instrumental detection limits (IDLs).

Multiplication by 3.14 of the mean deviation of the seven measurements is carried out to obtain MDLs at the 99% confidence level. The IDLs must be established prior to measuring the MDLs can be measured. IDLs were calculated by multiplying the standard deviation of 10 blank measurements by three.

Figure 5 illustrates the MDLs plotted beside the certified analyte concentrations in the four reference materials as analyzed in this study. (Note: not all analytes were held in the reference materials.)

Method detection limits (black) along with certified concentrations in four reference materials (shades of blue). Not all analytes were certified in the reference materials.

Figure 5. Method detection limits (black) along with certified concentrations in four reference materials (shades of blue). Not all analytes were certified in the reference materials. Image Credit: PerkinElmer, Inc.

The MDLs are considerably lower than the certified values, demonstrating the capacity of the methodology to measure low-level analytes in wastewaters with ease.

Linear Dynamic Range

Method 200.7 declares the linear dynamic range as the highest concentration, which recuperates within 10% of its true assigned value as measured against the calibration curve utilized for analysis.

All linear dynamic range measurements were conducted in multi-element solutions to ensure relevance to sample analyses, which are, in any case, multi-element solutions.

As Table 4 demonstrates, the highest concentration analyzed for most elements is the linear dynamic range. This is representative of the highest concentrations typically found in wastewaters, not the limit of the Avio 560 Max ICP-OES.

Table 4. Linear Dynamic Range. Source: PerkinElmer, Inc.

Elements Linear Range (mg/L)
Cd, Mn, Sr 30
Ba 40
Co, Cr, Ni, Sn, Ti, Tl 50
Be 70
Ag, Al, As, B, Cu, Fe, Li, Mg, Mo, P, Pb, Sb, Se, Si, V, Zn 100*
Ca, K, Na 500*

*= highest concentration evaluated

The linear range of the Avio 560 Max can be further extended, if necessary, by choosing less sensitive wavelengths, modifying the plasma viewing mode (i.e., axial/radial), switching the viewing height in the plasma to radial mode, varying the torch position, utilizing high-resolution mode and/or using a less sensitive sample introduction system.

Accuracy

With the base characteristics of the methodology determined, the accuracy was established via analysis of four reference materials, whose certified values are exhibited in Table 5. The analyte recoveries for each of the certified elements are shown in Figure 6.

Table 5. Certified Values in Wastewater Reference Materials (all units in mg/L) . Source: PerkinElmer, Inc.

Element Wastewater C Wastewater D Wastewater H WasteWATR
Ag 0.15 0.25 0.02 0.444
Al 0.5 1 0.1 2.91
As 0.15 0.25 0.1 0.151
B 0.5 1 0.25 1.33
Ba 0.5 1 0.1 1.5
Be 0.15 0.25 0.02 0.468
Cd 0.15 0.25 0.1 0.430
Co 0.5 1 0.5 0.147
Cr 0.5 1 0.5 0.302
Cu 0.5 1 0.5 0.526
Fe 0.5 1 0.25 2.25
Mn 0.5 1 0.1 0.636
Mo 0.5 1 0.1 0.578
Ni 0.5 1 0.5 1.21
Pb 0.5 1 0.5 1.05
Sb 0.15 0.25 0.2 0.693
Se 0.15 0.25 0.05 0.678
Sr 0.5 1 0.1 0.489
Tl 0.15 0.25 0.25 0.461
V 0.5 1 0.5 0.525
Zn 0.5 1 0.5 1.46

 

Figure 6. Certified analyte recoveries in four different wastewater reference materials.

Figure 6. Certified analyte recoveries in four different wastewater reference materials. Image Credit: PerkinElmer, Inc.

With all recoveries within 10% of their certified values, accuracy of the methodology is resolved.

Yet not all the elements detailed in Method 200.7 are certified in the reference materials.

The reference materials were contaminated with these elements, prior to digestion, at the concentrations exhibited in Table 6: their recoveries are shown in Figure 7. All recoveries are within 10%, further establishing the accuracy of the method.

Table 6. Spiked Concentrations of Non-Certified Elements in Wastewater CRMs. Source: PerkinElmer, Inc.

Analyte Concentration (mg/L)
Li, P, Si, Sn, Ti 0.75
Ca, K, Mg, Na 20

 

Recoveries of non-certified elements spiked into four wastewater CRMs.

Figure 7. Recoveries of non-certified elements spiked into four wastewater CRMs. Image Credit: PerkinElmer, Inc.

Stability

With the accuracy established, stability was proven by monitoring the recoveries of the IPC standards and measuring wastewater samples, which are run every ten samples.

Figure 8 illustrates how all IPCs over the 4-hour run recover within 10% of their certified values, with a sample-to-sample time of around 1 minute. This demonstrates the excellent stability of the system, despite the quick sample-to-sample time of ≈60 seconds.

IPC recoveries during a 4-hour run of wastewaters, with a sample-to-sample time of ˜ 60 seconds.

Figure 8. IPC recoveries during a 4-hour run of wastewaters, with a sample-to-sample time of ≈ 60 seconds. Image Credit: PerkinElmer, Inc.

Instrumental design considerations, such as the vertical torch and Flat Plate™ plasma technology, result in exceptional stability while making the most of the dual view capability and consistently moving between axial and radial plasma viewing modes for each sample.

Conclusion

Following the guidelines provided in U.S. EPA Method 200.7, this work illustrates the capacity of the Avio 560 Max ICP-OES to perform high-speed analysis – 60 sec sample-to-sample time – of wastewater.

With reliability, precision, robustness and stability exhibited during the analysis of reference materials and QC checks, the Avio 560 Max ICP-OES with integrated HTS offers a powerful solution for wastewater analysis while utilizing a total argon flow of 9 L/min.

The low argon consumption (9 L/min plasma flow) of the Avio 560 Max ICP-OES enables a swift return on investment for all laboratories, while the HTS considerably reduces sample uptake and washout times in contrast to traditional sample introduction, additionally increasing sample throughput.

References

1. Method 200.7, Revision 4.4: Determination of Metals and Trace Elements in Water and Wastes by Inductively Coupled Plasma- Atomic Emission Spectrometry”, United States Environmental Protection Agency, 1994.

2. “Analysis of Wastewaters Following U.S. EPA 200.7 using the Avio 550 Max ICP-OES”, Application Note, PerkinElmer Inc., 2020.

3. “High Throughput System for ICP-MS/OES,” Product Note, PerkinElmer Inc., 2020.

This information has been sourced, reviewed and adapted from materials provided by PerkinElmer.

For more information on this source, please visit PerkinElmer.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    PerkinElmer. (2023, May 23). How to Perform Rapid Water Analysis to Determine Trace Elements. AZoM. Retrieved on October 16, 2024 from https://www.azom.com/article.aspx?ArticleID=20073.

  • MLA

    PerkinElmer. "How to Perform Rapid Water Analysis to Determine Trace Elements". AZoM. 16 October 2024. <https://www.azom.com/article.aspx?ArticleID=20073>.

  • Chicago

    PerkinElmer. "How to Perform Rapid Water Analysis to Determine Trace Elements". AZoM. https://www.azom.com/article.aspx?ArticleID=20073. (accessed October 16, 2024).

  • Harvard

    PerkinElmer. 2023. How to Perform Rapid Water Analysis to Determine Trace Elements. AZoM, viewed 16 October 2024, https://www.azom.com/article.aspx?ArticleID=20073.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.