Reproducible Sample Preparation Using LeNeo® Instrument for ICP-OES Analysis of Fuel Oils

Fuel oil, a key by-product obtained from petroleum refining, is mainly utilized as fuel for cargo ships engines and ocean liners as well as in industrial plants and power plants.

Image Credits: shutterstock.com/ESB Professional

Petroleum refineries often use fluid catalytic cracking (FCC) process to optimize the distillation of high octane hydrocarbons[1]. The FCC process often involves some contaminants such as calcium, sodium, silicon, vanadium, iron, and nickel, which may be transferred into the fuel oil during the process.

It is crucial to know the contaminant concentration, as these elements become adhesive at certain temperatures, and can corrode reheater tube surfaces, combustion chambers, and other components in boilers and engines[2]. It is necessary to avoid corrosion at any cost as it leads to wear, thus leading to additional maintenance and replacement expenses.

This article demonstrates the use of borate fusion as a dissolution technique for inductively coupled plasma emission spectroscopy (ICP-OES) analysis to meet the accuracy, precision, and recovery prerequisites in standard methods such as ASTM D 5184[3], IP 377[4] and IP 501[5].

The LeNeo® fusion instrument (Figure 1) will facilitate this preparation step.

As outlined in the standards, a lithium metaborate with 1.5% lithium bromide integrated (LiM/1.5% LiBr) flux is employed in place of the lithium tetraborate blended with 10% lithium fluoride (LiT/LiF 90/10) flux to eliminate the loss of silicon caused by the reaction with fluoride, thereby increasing the solubility of the analytes present in the fuel oil.

The LeNeo® fusion instrument

Figure 1. The LeNeo® fusion instrument

Method

Apparatus and Instrument Conditions

Borate-based solutions were created using an automatic Claisse® LeNeo fusion instrument, which provides homogeneous heating through its preset fusion programs, superior insulation properties, and resistance-based electrical system. These features allow repeatable and reproducible fusion conditions to be achieved, in addition to a perfect retention of volatile elements.

The samples were ignited using a Fisher Scientific™ Isotemp™ muffle furnace. The data was collected using a PerkinElmer® Optima® 7300 DV ICP-OES spectrometer. Table 1 lists out the operating parameters used on the spectrometer.

Table 1. Optima 7300 DV operating parameters

Nebulizer GemTipTM Cross-Flow
Spray chamber Scott
Injector Alumina 2 mm
RF 1500 W
Argon flow Plasma: 16 L/min
Nebulizer: 0.8 L/min
Auxiliary: 0.4 L/min
Sample flow rate
1.0 mL/min

Global Sample Preparation Method

The following steps were involved in each sample preparation before being introduced into the fusion process:

  • Weigh the sample in a Pt/Au crucible
  • Burn the combustible material with an open flame
  • Ignite the carbon residue in a muffle furnace at 550°C overnight

After mixing the sample with 0.4g of LiM/1.5% LiBr flux in a Pt/Au crucible, it was fused using a fully automatic LeNeo instrument.

The melt was then automatically poured into 80mL of a 10% hydrochloric acid (HCl) solution, followed by magnetic agitation in order to produce a complete dissolution.

The LeNeo instrument took below 15 minutes to complete the entire fusion and dissolution process. The ensuing solutions were then made up to 100 mL in 10% HCl for further analyses on the PerkinElmer Optima 7300 DV ICP-OES.

Results

The high content of lithium in this matrix causes a suppression or boost in signals for some elements - this effect can be observed with many easily ionizable elements. This effect was counteracted by using matrix matching and internal standards. Sample dilution eliminated any clogging due to the high salt content of the matrix.

The LiM/1.5% LiBr matrix was used to prepare the calibration solutions of all elements, as described for the samples.

Calibration curves were generated using a concentration range close to the sample contentrations - all sample concentrations were within the range of the calibration. For each element, five points including the calibration blank were employed. Each curve had a correlation of over 0.999.

Method Detection Limits

Method detection limits (MDLs) were based on 10 repeated measurements on a set of diluted sample solutions. The standard deviation of the 10 duplicate measurements (S10) was multiplied by 3 to obtain the MDL (Table 2).

   MDL = 3 x S10

Method Concentration Range

The ICP-OES analysis method was validated for analyte determination within the concentration ranges listed in Table 2.

Table 2. Analytes of interest with selected wavelength, viewing mode, concentration range and method detection limit (MDL)

Elements Wavelengh (nm) View Concentration range (mg/kg) MDL(mg/kg)
Minimum Maximum
Al 396.153 Axial 5 150 0.01
Ca 396.847 Radial 3 100 0.3
Fe 259.939 Radial 2 100 0.03
Na 589.592 Radial 1 100 0.3
Ni 227.022 Axial 1 100 0.02
P 177.434 Axial 1 60 0.2
Si 251.611 Radial 10 250 0.2
V 270.093 Axial 1 400 0.04
Zn 202.548 Axial 1 70 0.02

Accuracy

Table 3 demonstrates the accuracy of the 10 duplicate measurements of four fuel oil samples.

These samples come from the round robin titled, “Committee D2 Proficiency Testing Program #6 Fuel Oil”. The mean values of each set of samples in the round robin were used to calculate the true values, and the accuracy values in Table 3 how close the results were to these true values.

Table 3. Accuracy measurements on 4 fuel oil samples

Elements F61409 (%) F61501 (%) F61505 (%) F61509 (%)
Al 99 91 89 89
Si 97 97 98 97

Recovery

The recovery values acquired in four similar matrixes are shown in Table 4. Recovery was estimated on five duplicates for each spiked sample.

Table 4. Recovery measurements on 4 fuel oil samples

Elements F61409 (%) F61501 (%) F61505 (%) F61509 (%)
Al 107 93 98 101
Ca 97 95 100 100
Fe 93 93 94 95
Na 98 93 101 106
Ni 95 101 97 101
P 103 97 98 97
Si 99 99 96 105
V 99 95 106 98
Zn 102 104 105 100

Precision

The standard deviation on 10 duplicate samples prepared and analyzed on different days was multiplied by 2.77[6] to obtain the experimental repeatability (SD10). Within the concentration range, the experimental repeatability should not go beyond the expected repeatability (SDstd) that was estimated as outlined by the standards[3,4,5] (Table 5).

Table 5. Repeatability measurements on 1 fuel oil sample

Elements 1
mg/kg
2
mg/kg
3
mg/kg
4
mg/kg
5
mg/kg
6
mg/kg
7
mg/kg
8
mg/kg
9
mg/kg
10
mg/kg
Average SD10
mg/kg
SDstd
mg/kg
Al* 3.0 3.2 3.3 3.2 3.6 3.4 3.0 3.0 3.3 3.5 3.2 0.6 0.2
Ca 9.6 10.1 9.7 9.7 10.3 9.7 10.1 10.2 10.4 9.6 9.9 0.8 1.7
Fe 27 29 28 28 28 27 28 27 28 28 28 2 4
Na 19 20 19 19 19 19 18 18 20 19 19 2 3
Ni 50 53 51 51 54 51 51 50 51 51 51 3 7
P 3.9 3.9 3.7 4.0 3.7 3.9 3.5 3.7 3.7 3.8 3.8 0.4 1.2
Si* 6.4 6.2 6.4 6.0 6.2 6.3 6.3 6.3 6.1 6.6 6.3 0.4 0.4
V 219 226 218 221 218 215 221 219 218 217 219 8 17
Zn 4.6 4.7 4.8 4.7 4.7 4.7 4.5 4.7 4.6 4.7 4.7 0.2 1.0

*Values outside the concentration range

Conclusion

The aforementioned results have demonstrated the suitability of the sample preparation process by fusion with a LiM/1.5% LiBr flux for the ICP-OES analysis of fuel oils.

The use of the LeNeo instrument for sample preparation is advantageous in many ways for all laboratories. It is simple, as it enables complete dissolution without acid heating. It is also fast, as it completes a fusion cycle within 15 minutes that includes the dissolution time.

With more than 89% accuracy and 93% to 107% recovery, the method shows its high reliability and validity. The good repeatability values obtained demonstrate the ability of the LeNeo instrument to provide a reproducible sample preparation, thereby meeting the reproducibility and repeatability as outlined by the standards ASTM D 5184, IP 377 and IP 501.

References

  1. "Octane-Enhancing Zeolitic FCC Catalysts: Scientific and Technical Aspects", J. Scherzer, Catal Rev, 1989. DOI: 10.1080/01614948909349934

  2. ISO 8217:2010 - Specifications of Marine Fuels, 4th Ed., International Organization for Standardization, 2010.

  3. ASTM D5184-12 - Standard Test Methods for Determination of Aluminum and Silicon in Fuel Oils by Ashing, Fusion, Inductively Coupled Plasma Atomic Emission Spectroscopy, and Atomic Absorption Spectroscopy, ASTM International, 2012.

  4. IP 377 - Determination of Aluminium and Silicon in Fuel Oils - Inductively Coupled Plasma Emission and Atomic Absorption Spectroscopy Methods, International Organization for Standardization, 2014.

  5. IP 501 - Determination of Aluminium, Silicon, Vanadium, Nickel, Iron, sodium, Calcium, Zinc and Phosphorous in Residual Fuel Oil by Ashing, Fusion and Inductively Coupled Plasma Emission Spectrometry, International Organization for Standardization, 2005.

  6. ASTM E117-08 - Standard Practice for Use of the Terms Precision and Bias in ASTM Test Methods. ASTM International, 2008.

Click here for more information about the LeNeo from Claisse

This information has been sourced, reviewed and adapted from materials provided by Claisse.

For more information on this source, please visit Claisse.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Malvern Panalytical. (2023, May 30). Reproducible Sample Preparation Using LeNeo® Instrument for ICP-OES Analysis of Fuel Oils. AZoM. Retrieved on December 02, 2024 from https://www.azom.com/article.aspx?ArticleID=12707.

  • MLA

    Malvern Panalytical. "Reproducible Sample Preparation Using LeNeo® Instrument for ICP-OES Analysis of Fuel Oils". AZoM. 02 December 2024. <https://www.azom.com/article.aspx?ArticleID=12707>.

  • Chicago

    Malvern Panalytical. "Reproducible Sample Preparation Using LeNeo® Instrument for ICP-OES Analysis of Fuel Oils". AZoM. https://www.azom.com/article.aspx?ArticleID=12707. (accessed December 02, 2024).

  • Harvard

    Malvern Panalytical. 2023. Reproducible Sample Preparation Using LeNeo® Instrument for ICP-OES Analysis of Fuel Oils. AZoM, viewed 02 December 2024, https://www.azom.com/article.aspx?ArticleID=12707.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.