Rapidly Assessing Wheat Gluten Vitality with the Rapid Visco Analyser (RVA)

The Rapid Visco Analyzer (RVA) is a cooking stirring viscometer that features ramped temperature and variable shear profiles. It has been optimized for testing a range of viscous properties.

Rapidly Assessing Wheat Gluten Vitality with the Rapid Visco Analyser (RVA)

Image Credit: PerkinElmer Food Safety and Quality

The instrument incorporates several international standard methods while offering full flexibility to utilize custom-made profiles.

The RVA’s robust combination of precision, speed, automation and flexibility make it an ideal solution for product development, quality assurance and process control. It is also well suited to the rapid assessment of wheat gluten vitality.

Description

Wheat gluten is primarily employed as an additive to improve leavened bread’s baking quality. It is also a popular choice for use in a range of other products, including pet foods.

Wheat gluten can be partially denatured by heat during commercial preparation, typically reducing its vitality. The RVA can be used to assess the functionality of wheat gluten via heating and cooling in aqueous ethanol.

RVA curves showing the large differences between vital and heat damaged gluten samples. Heat damaged samples lack the second peak.

Figure 1. RVA curves showing the large differences between vital and heat damaged gluten samples. Heat damaged samples lack the second peak. Image Credit: PerkinElmer Food Safety and Quality

Method

A 30-minute pasting profile was used for this study.

Sample Preparation

A total of 6.00 g gluten (12% mb) and 3.5 g ethanol (98%) were mixed. These were then stirred until the mixture was like breadcrumbs.

A total of 20.5 g distilled water was then added, bringing the total amount of water + flour to 26.5 g. The sample was then covered using a rubber stopper and vigorously shaken to disperse.

Profile

Source: PerkinElmer Food Safety and Quality

Time Type Value
00:00:00 Temp 50 °C
00:00:00 Speed 960 rpm
00:00:30 Speed 160 rpm
00:02:00 Temp 50 °C
00:06:30 Temp 85 °C
00:08:30 Temp 85 °C
00:24:00 Temp 20 °C
00:30:00 End  
Idle Temperature: 50 ± 1 °C
Time Between Readings: 4 s

 

Measure

Measurements were performed as follows:

  • P1V: Peak 1 viscosity (cP)
  • P2V: Peak 2 viscosity (cP)
  • P1T: Peak 1 temperature (°C)
  • P2T: Peak 2 temperature (°C)
  • T1V: Trough 1 viscosity (cP)
  • T2V: Trough 2 viscosity (cP)
  • T1T: Trough 1 temperature (°C)
  • T2T: Trough 2 temperature (°C)
  • FV: Final viscosity (cP)

The viscosity of the second peak can be understood to be the RVA Wheat Gluten Ethanol Index.

This information has been sourced, reviewed and adapted from materials provided by PerkinElmer Food Safety and Quality.

For more information on this source, please visit PerkinElmer Food Safety and Quality.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    PerkinElmer. (2023, May 22). Rapidly Assessing Wheat Gluten Vitality with the Rapid Visco Analyser (RVA). AZoM. Retrieved on April 27, 2024 from https://www.azom.com/article.aspx?ArticleID=20651.

  • MLA

    PerkinElmer. "Rapidly Assessing Wheat Gluten Vitality with the Rapid Visco Analyser (RVA)". AZoM. 27 April 2024. <https://www.azom.com/article.aspx?ArticleID=20651>.

  • Chicago

    PerkinElmer. "Rapidly Assessing Wheat Gluten Vitality with the Rapid Visco Analyser (RVA)". AZoM. https://www.azom.com/article.aspx?ArticleID=20651. (accessed April 27, 2024).

  • Harvard

    PerkinElmer. 2023. Rapidly Assessing Wheat Gluten Vitality with the Rapid Visco Analyser (RVA). AZoM, viewed 27 April 2024, https://www.azom.com/article.aspx?ArticleID=20651.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.