Particle Size Distribution Measurements on Materials

Developing battery technology is meeting with the demand for more portable devices, more power storage, and more time between recharges. Enhanced performance necessitates increased control over the materials used and their physical properties, including particle size distribution. In this study, the LA-950 laser diffraction particle size analyzer was utilized to execute particle size distribution measurements on a variety of the materials used in lithium ion batteries.

Introduction

Lithium batteries include disposable (primary), rechargeable (secondary), and rechargeable lithium titanate batteries. Disposable lithium batteries use lithium metal or lithium compounds as the anode. Rechargeable lithium ion batteries include an intercalated lithium compound as the electrode material. A lithium titanate battery is an adapted lithium-ion battery that uses lithium titanate nanocrystals on the surface on the anode.

Some cathode anode materials used in lithium batteries are:

Cathode

  • Lithium cobalt oxide LiCoO2
  • Lithium nickel oxide LiNiO2
  • Lithium manganese oxide LiMn2O4
  • Lithium iron phosphate LiFePO4

Anode

  • Carbon C
  • Lithium Li
  • Lithium titanate Li2TiO3

The particle size distribution (PSD) of the materials used in the production of batteries is tested both during R&D and in product acceptance QC, as PSD is generally a material specification. Particle size affects capacity and coulomb efficiency. A smaller PSD leads to a bigger specific surface area, changing critical battery properties and causing variations in the size of the voids between electrode particles, reducing battery capacity.

Experimental

Lithium Cobalt Oxide

Lithium cobalt oxide (LiCoO2) has been the standard cathode material used for secondary Li-ion batteries for over three decades. Five distinct lots of lithium cobalt oxide powder to be used as cathode material were examined on the LA-950 Particle Size Analyzer for a standard pass/fail test to determine if the material met the incoming material specification. The powder was dispersed in water containing 0.2% sodium hexametaphosphate, with the results displayed in Figure 1 and Table 1.

Five lots of LiCoO2 powder

Figure 1. Five lots of LiCoO2 powder

Table 1. Five lots of LiCoO2 powder

Lot Median (m)
No. 1 11.3
No. 2 11.8
No. 3 12.2
No. 4 12.5
No. 5 11.9

 

Lithium Manganese Oxide

While developing the method for dispersing lithium manganese oxide (LiMn2O4), the effect of using ultrasound was explored. The powder was dispersed in DI water containing 0.2% sodium hexametaphosphate. The sample was then studied both without ultrasound, and with one, three, and five minutes of ultrasound.

The effect of ultrasound on LiMn2O4

Figure 2. The effect of ultrasound on LiMn2O4

Table 2. The effect of ultrasound on LiMn2O4

Ultrasound Median (m) Color
None 10.46 blue
1 min 10.14 light green
3 min 9.67 red
5 min 9.42 dark green

 

The results (Figure 2, Table 2) indicate that three minutes was the ideal time. To prove reproducibility, the method was repeated on the same sample for five days (Figure 3, Table 3).

Method validation testing on multiple days of LiMn2O4

Figure 3. Method validation testing on multiple days of LiMn2O4

Table 3. Method validation testing on multiple days of LiMn2O4

  Median (m)
Day 1 9.75
Day 2 9.52
Day 3 9.18
Day 4 9.49
Day 5 9.5
Mean 9.49
COV 1.90%

 

Lithium Titanate

Lithium titanate (Li2TiO3) is frequently applied as the anode in fast charging lithium titanate batteries. The effect of ultrasound on dispersing lithium titanate powder was investigated. The powder was dispersed in DI water containing 0.2% sodium hexametaphosphate. The sample was analyzed without and with three minutes of ultrasound. The results (Figure 4) indicate sample dispersion did not improve with use of ultrasound, so its use was discarded.

Li2TiO3 with (green) and without (blue) ultrasound

Figure 4. Li2TiO3 with (green) and without (blue) ultrasound

Two samples from different suppliers were then compared by measuring the PSD of both using the LA-950. The powders were dispersed in DI water containing 0.2% phosphoric acid and 0.2% sodium hexametaphosphate. The comparative results for Sample A (mean = 6.33 µm) and Sample B (mean= 16.7 µm) are shown in Figure 5.

Li2TiO3 sample A (red) vs. Sample B (green) Reproducibility: Lithium manganese oxide/Lithium titanate

Figure 5. Li2TiO3 sample A (red) vs. Sample B (green) Reproducibility: Lithium manganese oxide/Lithium titanate

Two samples (LiMn2O4 and Li2TiO3 ) were analyzed 10 times to quantify the reproducibility of the LA-950 (Figure 6). The COV values are very low, indicating that the method is effective, and demonstrating the high performance level of the LA-950 particle size analyzer.

Reproducibility over 10 results for LiMn2O4 and Li2TiO3

Sample LiMn2O4 Li2TiO3
R1 9.75 16.70
R2 9.93 16.6
R3 9.75 16.7
R4 9.66 16.6
R5 9.83 16.7
R6 9.78 16.6
R7 9.76 16.8
R8 9.75 16.8
R9 9.79 16.8
R10 9.6 16.9
Ave 9.76 16.7
COV 0.90% 0.51%

 

Figure 6. Reproducibility over 10 results for LiMn2O4 and Li2TiO3

Instrument to Instrument Agreement

Test results from different laboratories need to be compared by companies that have multiple locations. Samples of lithium manganese oxide and lithium titanate were studied on two different LA-950 systems in order to quantify the level of agreement between instruments (Figure 7).

Instrument to instrument agreement for LiMn2O4 and Li2TiO3 on two different LA-950 systems

  950 1 950 2 Δ
LiMnO 9.75 9.64 0.1
LiTi 16.7 16.9 0.2

 

Figure 7. Instrument to instrument agreement for LiMn2O4 and Li2TiO3 on two different LA-950 systems

Conclusions

When measuring the PSD of many battery materials, the LA-950 demonstrated excellent reproducibility and agreement between systems. This performance has been recognized by several manufacturers of battery materials, leading to numerous sales to suppliers and users around the world.

Acknowledgements

Many thanks to the HORIBA Applications Center in Kyoto for generating the excellent data shown in this application note.

This information has been sourced, reviewed and adapted from materials provided by HORIBA.

For more information on this source, please visit HORIBA.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    HORIBA. (2024, March 19). Particle Size Distribution Measurements on Materials. AZoM. Retrieved on April 20, 2024 from https://www.azom.com/article.aspx?ArticleID=16133.

  • MLA

    HORIBA. "Particle Size Distribution Measurements on Materials". AZoM. 20 April 2024. <https://www.azom.com/article.aspx?ArticleID=16133>.

  • Chicago

    HORIBA. "Particle Size Distribution Measurements on Materials". AZoM. https://www.azom.com/article.aspx?ArticleID=16133. (accessed April 20, 2024).

  • Harvard

    HORIBA. 2024. Particle Size Distribution Measurements on Materials. AZoM, viewed 20 April 2024, https://www.azom.com/article.aspx?ArticleID=16133.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Your comment type
Submit

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.